VISIR trending system: HEALTH CHECK report
Last update: 2012-05-24T02:05:08 (UT)
| now: 2012-05-24T23:04:33 (UT)
HELP USERS-GUIDE MORE
same group: read-noise read-noise12 dark flat-flux nbadpix det-temp
General news: 2012-05-24: The HC monitors for OMEGACAM, VIRCAM and HAWKI are back. Population with current data is in progress.
VISIR news: 2012-05-09: VISIR upgrade began; instrument is scheduled back in operations on Sep.9th, 2012
Report news:  
DATE*: [?] 2012-05-17
no data | NR
2012-05-18
no data | NR
2012-05-19
no data | NR
2012-05-20
no data | NR
2012-05-21
no data | NR
2012-05-22
no data | NR
2012-05-23
no data | NR
2012-05-24
no data | NR
P...
Product quality:
no products no products no products no products no products no products no products no products
  last OPSLOG data:
2012-05-08*

last files ...
monitor

scores&comments | history ... | plot tutorial ... | contact                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
OPS?
**
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB yes (0) MEAN -28058.00195 ADU VAL-28080,-27980 72 darkmed this | last_yr | all mean DARK level of the imaging detector
2 QC1DB yes (0) MEAN -30996.8418 ADU VAL-31200,-30900 141 darkmed this | last_yr | all mean DARK level of the spectroscopic detector
*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: to indicate that OPSLOG data are included
Plot 1
scores: 2|
data source:visir_dark_img
(QC1 database)
dataset:darkmed
mean:-28058.00195ADU
fixed thresholds:-28080.0...-27980.0ADU
N_data plotted:72
[click on plot for closeup]
Plot 2
scores: 4|
data source:visir_dark_spc
(QC1 database)
dataset:darkmed
mean:-30996.8418ADU
fixed thresholds:-31200.0...-30900.0ADU
N_data plotted:141
[click on plot for closeup]
This plot

Trending plots show dark level of the imaging and spectroscopic detectors of VISIR.  


General information

Click on any of the plots to see a close-up version.

If applicable, the latest values from Paranal ops logs are plotted as blue dots. More information about those values, including a download option, is available in the OPS column.

The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up