XSHOOTER trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2025-12-16T22:36:00 (UT)
  | now: 2025-12-17T07:01:06 (UT) muc02 QC pipeline: xshoo-3.6.5a (installed 2023-09-18)
same group: bias_AGC dark_AGC detmon_AGC
close window history: 2014 2015 2016 2017 2018 2019 2020 2021 2022 2023 2024   2025   HEALTH
  ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 1.78e+03 ADU 20%± 355 46 median_master this | last_yr | all  
2 QC1DB MEDIAN 0.975 ADU 80%± 0.78 46 struct_x1 this | last_yr | all  
3 QC1DB MEDIAN 4.44 ADU 60%± 2.67 46 ron this | last_yr | all  
4 QC1DB MEDIAN 0.593 ADU 60%± 0.356 46 struct_y1 this | last_yr | all  
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 7|
data source:xshooter_bias
(QC1 database)
dataset:median_master
median:1.78e+03ADU
rel. thresholds: ±20%1420.267578...2130.401367ADU
N_data plotted:46
[click plot for closeup]
Plot 3
scores: 7|
data source:xshooter_bias
(QC1 database)
dataset:ron
median:4.44ADU
rel. thresholds: ±60%1.7775464...7.1101856ADU
N_data plotted:46
[click plot for closeup]
Plot 2
scores: not implemented
data source:xshooter_bias
(QC1 database)
dataset:struct_x1
median:0.975ADU
rel. thresholds: ±80%0.1950617...1.7555553ADU
N_data plotted:46
[click plot for closeup]
Plot 4
scores: not implemented
data source:xshooter_bias
(QC1 database)
dataset:struct_y1
median:0.593ADU
rel. thresholds: ±60%0.2370958...0.9483832ADU
N_data plotted:46
[click plot for closeup]
This plot

This plots shows detector related QC parameters derived from BIAS health check calibrations of the xshooter acquisition and guiding camera (AGC). It is a 562 x 528 pixel technical CCD. For the definition of the QC parameters, please see the corresponding section of the UVB-arm and VIS-arm science detector. On 2014-01-09 the detector operations mode was switched from DET.OUT1.GAININD = 8 to DET.OUT1.GAININD = 4.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up