Health Check monitor


Agora | CAL | HC | refs | QC
HealthCheck Monitor
    HOME | UsersGuide
FORS1:
zeropoints: frame
Other HC:
QC links:
FORS1 trending system: HEALTH CHECK plot
Last update: 2008-05-20T22:48:55 (UT)
now: 2008-12-04T07:15:36 (UT)
INFO USERS-GUIDE MORE
same group: U_BESS B_BESS V_BESS R_BESS I_BESS U_HIGH B_HIGH V_HIGH G_HIGH
history ... | plot tutorial ... [info]
last OPSLOG data:
2008-04-19
Plot: 1 2 3 4   latest_files
N_data: number of recent OPSLOG entries plotted here
N_data: 0 0 0 0 availability
Print: Information and research: News:
- png file - advanced studies: QC1 browser | QC1 plotter
 
Plot
[info]
Symbol
[info]
Source* OPS?** Average [info] Thresholds [info] N_data Data downloads Remarks
method value unit method value
1 QC1DB yes (0) none   mag none   43 this | last_yr | all Frame zeropoint from e-/sec (more than 1 star; Garching)
1 QC1DB yes (0) none   mag none   4 this | last_yr | all Frame zeropoint from e-/sec (1 star only; Garching)
1 QC1DB yes (0) MEDIAN 28.18650 mag 3SIG± 0.08597 40 this | last_yr | all Frame zeropoint from e-/sec (photometric nights only; Garching)
2 QC1DB yes (0) MEDIAN 0.01383 mag 3SIG± 0.02363 47 this | last_yr | all error of frame zeropoint from e-/sec (Garching)
3 QC1DB yes (0) none   mag none   38 this | last_yr | all Frame zeropoint from e-/sec (more than 1 star; Garching)
3 QC1DB yes (0) none   mag none   0 this | last_yr | all Frame zeropoint from e-/sec (1 star only; Garching)
3 QC1DB yes (0) MEDIAN 28.12524 mag 3SIG± 0.07324 36 this | last_yr | all Frame zeropoint from e-/sec (photometric nights only; Garching)
4 QC1DB yes (0) MEDIAN 0.01145 mag 3SIG± 0.02124 38 this | last_yr | all error of frame zeropoint from e-/sec (Garching)
*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: to indicate if OPSLOG data are included
This plot

Trending is done only for the SR collimator. For the old CCD (until 2007-03-25) only 1x1 binning data are trended. For the new CCDs (since 2007-04-01) only 2x2 binning data are trended. Photometric nights are selected from the Absorption Monitor. Zeropoints are per frame for flux in e-/sec, not in ADU. Do not use these zeropoints for science data. They are reduced assuming a standard extinction value, which may not be correct for an individual night!  


General information

Click on any of the plots to see a close-up version.

If applicable, the latest values from Paranal ops logs are plotted as blue dots. More information about those values, including a download option, is available in the OPSLOG section on top of the page.

The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up