SINFONI trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2014-11-28T05:04:05 (UT)
  | now: 2014-11-28T11:20:20 (UT) muc04 QC pipeline: sinfo-2.5.4 (installed 2014-10-20)
same group: slit_dist_ALL slit_dist_JHK distortion_shift   distortion_coef00 coef01 coef10 coef11 coef20 coef02 coef21
close window history: 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014   HEALTH FULL
  ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
OPS?
**
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 oQC1DB no MEDIAN -13.2 pixels VAL-17.15,-2.0 235 qc_xshift_ul this | last_yr | all distortion in upper left quadrant of detector for ALL bands [pixels]
(this FULL data set is scored)
1 QC1DB no none   pixels none  51 qc_xshift_ul this | last_yr | all J-band distortion: upper left quadrant of detector [pixels]
1 QC1DB no none   pixels none  34 qc_xshift_ul this | last_yr | all H-band distortion: upper left quadrant of detector [pixels]
1 QC1DB no none   pixels none  59 qc_xshift_ul this | last_yr | all K-band distortion: upper left quadrant of detector [pixels]
1 QC1DB no none   pixels none  91 qc_xshift_ul this | last_yr | all H+K-band distortion: upper left quadrant of detector [pixels]

2 oQC1DB no MEDIAN -7.64 pixels VAL-10.0,4.0 235 qc_xshift_ur this | last_yr | all distortion in upper right quadrant of detector for ALL bands [pixels]
(this FULL data set is scored)
2 QC1DB no none   pixels none  51 qc_xshift_ur this | last_yr | all J-band distortion: upper right quadrant of detector [pixels]
2 QC1DB no none   pixels none  34 qc_xshift_ur this | last_yr | all H-band distortion: upper right quadrant of detector [pixels]
2 QC1DB no none   pixels none  59 qc_xshift_ur this | last_yr | all K-band distortion: upper right quadrant of detector [pixels]
2 QC1DB no none   pixels none  91 qc_xshift_ur this | last_yr | all H+K-band distortion: upper right quadrant of detector [pixels]

3 oQC1DB no MEDIAN -9.93 pixels VAL-13.0,1.0 235 qc_xshift_cc this | last_yr | all distortion in center of the detector for ALL bands [pixels]
(this FULL data set is scored)
3 QC1DB no none   pixels none  51 qc_xshift_cc this | last_yr | all J-band distortion: center of detector [pixels]
3 QC1DB no none   pixels none  34 qc_xshift_cc this | last_yr | all H-band distortion: center of detector [pixels]
3 QC1DB no none   pixels none  59 qc_xshift_cc this | last_yr | all K-band distortion: center of detector [pixels]
3 QC1DB no none   pixels none  91 qc_xshift_cc this | last_yr | all H+K-band distortion: center of detector [pixels]

4 oQC1DB no MEDIAN -13.9 pixels VAL-17.25,-2.0 235 qc_xshift_ll this | last_yr | all distortion in lower left quadrant of detector for ALL bands [pixels]
(this FULL data set is scored)
4 QC1DB no none   pixels none  51 qc_xshift_ll this | last_yr | all J-band distortion: lower left quadrant of detector [pixels]
4 QC1DB no none   pixels none  34 qc_xshift_ll this | last_yr | all H-band distortion: lower left quadrant of detector [pixels]
4 QC1DB no none   pixels none  59 qc_xshift_ll this | last_yr | all K-band distortion: lower left quadrant of detector [pixels]
4 QC1DB no none   pixels none  91 qc_xshift_ll this | last_yr | all H+K-band distortion: lower left quadrant of detector [pixels]

5 oQC1DB no MEDIAN -6.11 pixels VAL-10.0,6.0 235 qc_xshift_lr this | last_yr | all distortion in lower right quadrant of detector for ALL bands [pixels]
(this FULL data set is scored)
5 QC1DB no none   pixels none  51 qc_xshift_lr this | last_yr | all J-band distortion: lower right quadrant of detector [pixels]
5 QC1DB no none   pixels none  34 qc_xshift_lr this | last_yr | all H-band distortion: lower right quadrant of detector [pixels]
5 QC1DB no none   pixels none  59 qc_xshift_lr this | last_yr | all K-band distortion: lower right quadrant of detector [pixels]
5 QC1DB no none   pixels none  91 qc_xshift_lr this | last_yr | all H+K-band distortion: lower right quadrant of detector [pixels]

*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: OPSLOG data are included
Plot 1
scores: 3|
data source:sinfoni_dist
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_xshift_ulo
median:-13.2pixels
fixed thresholds:-17.15...-2.0pixels
N_data plotted:235
[click on plot for closeup]
Plot 2
scores: 3|
data source:sinfoni_dist
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_xshift_uro
median:-7.64pixels
fixed thresholds:-10.0...4.0pixels
N_data plotted:235
[click on plot for closeup]
Plot 3
scores: 3|
data source:sinfoni_dist
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_xshift_cco
median:-9.93pixels
fixed thresholds:-13.0...1.0pixels
N_data plotted:235
[click on plot for closeup]
Plot 4
scores: 3|
data source:sinfoni_dist
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_xshift_llo
median:-13.9pixels
fixed thresholds:-17.25...-2.0pixels
N_data plotted:235
[click on plot for closeup]
Plot 5
scores: 3|
data source:sinfoni_dist
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_xshift_lro
median:-6.11pixels
fixed thresholds:-10.0...6.0pixels
N_data plotted:235
[click on plot for closeup]
This plot

These are the trending plots for the SINFONI QC parameters derived from DISTORTION frames.

Distortion frames are obtained for the J, H, K, and H+K bandpasses, but only for the 250 mas camera.

 


General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up