[ ESO ]
CRIRES Quality Control:
spectro-photometric standards

QUALITY CONTROL
  HOME
CRIRES QC
Trending & QC1
   std
Pipeline
Service Mode
Data Management
QC links:
[best viewing with style sheets and javascript enabled]
 
QC PLOTS
  CURRENT HISTORY
STD efficiency
QC1 database (advanced users): browse | plot
   Click on CURRENT to see the current trending (Health Check).
   Click on HISTORY to see the historical evolution of the trending.


Telluric and spectro-photometric standard stars are chiefly observed on request of the Service Mode or Visitor Mode observer. For health check purposes, spectro-photometric standards are frequently measured for a small number of settings: 1087.3, 1504.8, 2205.0, 3611.0, and 5050.4 nm.

Standard stars are reduced in the same way as science exposures (see the pipeline recipe crires_spec_jitter). For spectro-photometric standards, the instrumental throughput, sensitivity, and conversion factor per wavelength are calculated in addition.

top Spectro-photometric standards

QC1 parameters for CRIRES standard star exposures are inserted into the QC1 database only for spectro-photometric standards (and not for telluric standards). They are calculated for each detector array separately.

QC1 parameters

parameter QC1 database: table, name procedure
median conversion factor crires_std, med_conv median of conversion factor (in ADU/sec/Jy); the ratio of the extracted flux to the tabulated flux of the standard star
median troughput crires_std, med_through median of the throughput per wavelength (in e-/photon); the number of detected electrons per photon
median sensitivity crires_std, med_sens median of the sensitivity (in Jy/10sigma/hour)
spectral position crires_std, spec_pos median of the spectral position (in Y direction) on the detector

Trending

Throughput is trended for the five health check settings.

 


 
[ESO][Index][Search][Help][News]