mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
DPI Flats & Lamps
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
SPHERE trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-04-26T04:43:27 (UT)
  | now: 2024-04-26T12:36:36 (UT) muc03 QC pipeline: spher-0.52.0 (installed 2024-03-22)
same group: DPI_HIGH_FLAT DPI_BACKG
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
SPHERE news: Note that the current Health Check monitor runs with an incomplete pipeline and an incomplete set of QC1 parameters. QCG can currently not guarantee completeness and significance at the usual level.
Report news:
DATE*: [?] 2024-04-19
report | NLT
2024-04-20
report | NLT
2024-04-21
report | NLT
2024-04-22
report | NLT
2024-04-23
no data | NLT
2024-04-24
no data | NLT
2024-04-25
no data | NLT
r...
Raw CAL displays:
raw raw raw raw no CAL raw files no CAL raw files no CAL raw files
P...
CAL product quality:
products products products products no CAL products no CAL products no CAL products
 
transfer   ngas [?]

scores&comments | history ... | plot tutorial ... | contact | HC_docu... | quality events...               [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 8835.21 ADU VAL5000,25000 3 qc_adu_left_4 this | last_yr | all Y-band left beam
2 QC1DB MEDIAN 8744.88 ADU VAL5000,25000 3 qc_adu_right_4 this | last_yr | all Y-band right beam
3 QC1DB MEDIAN 10135.6 ADU VAL5000,25000 7 qc_adu_left_4 this | last_yr | all J-band left beam
4 QC1DB MEDIAN 11260.3 ADU VAL5000,25000 7 qc_adu_right_4 this | last_yr | all J-band right beam
5 QC1DB MEDIAN 11941.8 ADU VAL5000,25000 24 qc_adu_left_4 this | last_yr | all H-band left beam
6 QC1DB MEDIAN 11638.7 ADU VAL5000,25000 24 qc_adu_right_4 this | last_yr | all H-band right beam
7 QC1DB MEDIAN 18173.4 ADU VAL5000,25000 5 qc_adu_left_4 this | last_yr | all K-band left beam
8 QC1DB MEDIAN 14754.5 ADU VAL5000,25000 5 qc_adu_right_4 this | last_yr | all K-band right beam
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: not implemented
data source:sphere_irdis_flat
(QC1 database)
dataset:qc_adu_left_4
median:8835.21ADU
fixed thresholds:5000...25000ADU
N_data plotted:3
[click plot for closeup]
Plot 2
scores: not implemented
data source:sphere_irdis_flat
(QC1 database)
dataset:qc_adu_right_4
median:8744.88ADU
fixed thresholds:5000...25000ADU
N_data plotted:3
[click plot for closeup]
Plot 3
scores: not implemented
data source:sphere_irdis_flat
(QC1 database)
dataset:qc_adu_left_4
median:10135.6ADU
fixed thresholds:5000...25000ADU
N_data plotted:7
[click plot for closeup]
Plot 4
scores: not implemented
data source:sphere_irdis_flat
(QC1 database)
dataset:qc_adu_right_4
median:11260.3ADU
fixed thresholds:5000...25000ADU
N_data plotted:7
[click plot for closeup]
Plot 5
scores: not implemented
data source:sphere_irdis_flat
(QC1 database)
dataset:qc_adu_left_4
median:11941.8ADU
fixed thresholds:5000...25000ADU
N_data plotted:24
[click plot for closeup]
Plot 6
scores: not implemented
data source:sphere_irdis_flat
(QC1 database)
dataset:qc_adu_right_4
median:11638.7ADU
fixed thresholds:5000...25000ADU
N_data plotted:24
[click plot for closeup]
Plot 7
scores: not implemented
data source:sphere_irdis_flat
(QC1 database)
dataset:qc_adu_left_4
median:18173.4ADU
fixed thresholds:5000...25000ADU
N_data plotted:5
[click plot for closeup]
Plot 8
scores: not implemented
data source:sphere_irdis_flat
(QC1 database)
dataset:qc_adu_right_4
median:14754.5ADU
fixed thresholds:5000...25000ADU
N_data plotted:5
[click plot for closeup]
This plot

This plot shows the median counts of the last raw flat frame of a sequence of raw flats with increasing DIT. The QC parameter is monitored for the four broad band filters and for both channels.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up