[ ESO ]
VIRCAM Quality Control:
Darkcurrent

CAL | HC | refs | QC
QUALITY CONTROL
    HOME
VIRCAM QC
Trending & QC1
   darkcurrent
Pipeline
QC links:
darkcurrent
 
QC PLOTS
  CURRENT HISTORY
darkcurrent
QC1 database (advanced users): browse | plot
   Click on CURRENT to see the current trending (Health Check).
   Click on HISTORY to see the historical evolution of the trending.


The darkcurrent calibration is a technical calibration. It is not used to calibrate other calibrations or science data. The darkcurrent template acquires a sequence of dark frames with different DITs. The pipeline recipe fits the relation darkcurrent( DIT ). While the dark level retrieved from the dark recipe gives the dark counts in ADU for a master dark of a given DIT, the darkcurrent gives the current in ADU/sec in a more accurate manner.

top dark current

QC1 parameters

parameter QC1 database: table, name procedure

darkcurrent

vircam_current, qc_darkcurrent - The QC.DARKCURRENT gives the dark current in ADU/sec.

QC1 report (from 2009-10)

click in the image to see dark current fit of all detectors


Each detector has its own dark current. There following map shows the values in ADU/sec (June 2010):

chip13
chip 14
chip 15
chip 16
0.405
0.221
0.030
0.052
chip 9
chip 10
chip 11
chip 12
0.065
0.038
0.059
0.053
chip 5
chip 6
chip 7
chip 8
0.066
0.079
0.090
0.097
chip 1
chip 2
chip 3
chip 4

0.296

3.000
0.476
0.355

red labels mark detectors with relatively high dark current, green labels mark detectors with relatively low dark current.

 

 


 
 
[ESO][Index][Search][Help][News]