VISIR
CALIBRATION DATA: GENERAL | Calibration
data for VISIR are taken for two purposes:
- to assess the instrument status
- to reduce science data
VISIR calibration data are taken at night (e.g. standard star
observations) and as daytime calibrations (flatfield). On request of
the user, also nighttime calibrations can be obtained as part of the
Observing Block which generates the SCIENCE data ("attached
calibrations").
All calibration data taken in VISIR are processed by QC Garching to
obtain calibration solutions. These solutions come as
- calibration product fits file frames (e.g. standard star imaging frames)
- calibration product fits tables (e.g. spectrum extracted in a form of a table)
|
|
|
TYPES OF RAW CALIBRATION
DATA |
|
| FITS header examples |
| frame |
.txt |
| IMG FLAT |
|
| SPEC FLAT |
|
| IMG STD |  |
| SPEC STD |  |
|
The following types of raw calibration data exist for the VISIR:
| frame | DPR
TYPE | Purpose |
typical number taken per night and setting |
| IMG FLAT | FLAT | fixed pattern noise, gain, bad pixel map | 7 x each filter/lens combination |
| SPEC FLAT | FLAT | fixed
pattern noise, gain, bad pixel map | 7 x each instr. combination |
| IMG STANDARD STAR | STD | standard
stars for imaging | varies |
| SPEC STANDARD STAR | STD | standard
stars for telluric features | varies |
| |
![[top]](../img/arr1red-up.gif) | CALIBRATION
PRODUCTS |
|
|
Recipes and products
The
following VISIR calibration products are created from raw calibration files
(click here for a complete overview of calibration
files and recipes):
| raw frame(s) | template/recipe | description |
products, PRO.CATG |
| IMG FLAT |
VISIR_img_cal_ExtSrcFlats visir_img_ff | creates
a master flatfield for the imaging detector from a set of 7 input raw frames taken at different temp. of the WarmCalibrationUnit |
|
| SPEC FLAT |
VISIR_spec_cal_ExtSrcFlats visir_spc_ff | creates
a master flatfield for the spectroscopic detector from a set of 7 input raw frames taken at different temp. of the WarmCalibrationUnit |
|
| IMG STD | VISIR_img_cal_AutoChopNod visir_img_phot | determines observing conditions, instrumental respons, senistivity, background, etc. | IMG_PHOT_COMBINED IMG_PHOT_CONTRIB_MAP |
| SPEC STD | VISIR_spec_cal_LRAutoChopNod visir_spc_phot | telluric standard, determines observing conditions, instrumental respons, senistivity, etc. | SPC_PHOT_COMBINED SPC_PHOT_TAB SPC_PHOT_WEIGHT |
| Naming
scheme After production and certification, the calibration products
are renamed after a scheme which includes type,
creation data and relevant VISIR setting. . |
![[top]](../img/arr1red-up.gif) | CALIBRATION
CASCADE |
|
|
At present, all sets of VISIR calibration frames are reduced independently,
i.e. there is no calibration cascade. In the future the flatfield correction
may be included. It's still under investigation.
|
![[top]](../img/arr1red-up.gif) | OTHER
CALIBRATION TABLES |
|
|
Static calibrations include a reference arc line table and a quantum efficiency
map
|
![[top]](../img/arr1red-up.gif) | ACTUAL
PROBLEMS AND ISSUES |
| |
VISIR Pipeline problems are documented under on the ServiceMode
page. |