2|c| Parameters: CALIBRATE command (See also next table)
Parameter
Description
INSTR
Defines the spectrograph (e.g. CASPEC, EFOSC).
GRATING
Defines the grating or instrument configuration.
CCD
Defines the CCD.
CCDBIN
Binning factor along X and Y. This parameter has impact on the
wavelength calibration.
SCAN
Limits of the scanned area of the CCD. This parameter has impact on
order definition (method Hough) and background estimates.
ORDREF
Order reference frame
DEFMTD
Order definition method (STD, COM, HOUGH).
DEFPOL
Degree of the bivariate polynomial used in the definition of
the orders, see Section 7.2.
WIDTH1
Defines the width in pixels of the orders perpendicular to
the dispersion direction. It is used to detect the orders as
described in Section 7.2 (methods STD, COM).
Typical values are between 2 and 15, depending on the observing slit.
THRES1
Defines the threshold to detect orders on the order reference image as
described in Section 7.2 (methods STD, COM).
SLOPE
Mean slope of the orders in pixel space (methods STD, COM).
2|c| Parameters: CALIBRATE command (Continued)
Parameter
Description
NBORDI
Number of orders to be detected (method HOUGH).
WIDTHI
This parameter is used to remove the traces of the orders in Hough
space during the detection. Its minimum value is the half-width in
pixels of the orders perpendicular to the dispersion direction.
Its maximum value is the interorder distance minus the half-width
of the orders (method HOUGH).
THRESI
Defines the threshold for the order detection (equivalent to THRES1)
WLC
Raw wavelength calibration image.
LINCAT
System table with comparison lines.
EXTMTD
Extraction method (AVERAGE, LINEAR, OPTIMAL).
SLIT
Width of the extraction slit in pixels.
OFFSET
Offset of the extraction slit.
The offset is given in pixels, positive values are above the center
and negative offsets are below the center of the orders as shown in
the image monitor (command LOAD/ECHELLE).
SEAMTD
Searching method (GAUSSIAN, GRAVITY). This parameter controls the
definition of the center of the calibration lines.
WIDTH2
Defines the width of the calibration lines in pixels.
THRES2
Defines the threshold above the local background to detect comparison
lines. The actual value should be chosen to detect about
10 to 20 lines per order.
WLCMTD
Wavelength calibration method (PAIR, ANGLE, TWO-D, GUESS, RESTART, ORDER).
This parameter controls the execution of IDENTIFY/ECHELLE.
WLCOPT
Dispersion relation option (1D, 2D)
GUESS
Optional reference to a previous session with the same
instrument configuration, to be used as a
guess for the dispersion coefficients (required if WLCMTD=GUESS).
WLCVISU
Visualisation of intermediate results (YES/NO).
DC
Is the degree of the polynomial used in the dispersion relation,
see Section 7.6.
TOL
Defines a tolerance window for the last step of single-order
computation of the dispersion relations.
WLCLOOP
Three parameters to control the iterative identification of lines
by improvement of a global solution.
WLCNITER
Two parameters for the minimum and maximum number of iterations
of the identification loop.
2|c| Parameters: FLAT command
Parameter
Description
FFOPT
Flat-Fielding option (YES/NO). If the value NO is assigned
to this paramter, no Flat-Field correction is performed
FLAT
Raw flat field image.
CORRECT
Flat field correction image.
BLAZE
Smoothed flat field used in the normalisation of the flat field.
BKGMTD
Background estimation method (POLY, SPLINE, SMOOTH).
BKGVISU
Visualisation of intermediate results (YES/NO).
BKGSTEP
Step in pixels of the grid of background reference positions.
BKGPOL
Degree of the bivariate polynomial for background fiting (method POLY).
BKGRAD
Radius (radx, rady) in pixels of the window for local estimate of the
background. The window is parallel to the orders (radx is measured along
the orders, rady is parallel to the columns). (method SPLINE)
BKGDEG
Degree of spline polynomials (method SPLINE).
BKGSMO
Smoothing factor for smoothing spline interpolation. See Section
7.4 (method SPLINE).
EXTMTD
(See definition in Table 7.7)
SLIT
(See definition in Table 7.7)
OFFSET
(See definition in Table 7.7)
SAMPLE
Defines the wavelength step of the spectrum sampled in wavelengths.
It can either be a real number indicating the step in units
of Ångstrøm or a reference file corresponding to an image sampled
in the space order-wavelength.
2|c| Parameters: FILTER command
Parameter
Description
BKGxxx
All background parameters : see definitions above.
MEDFILT
widx,widy,no_iter. See help of command FILTER/ECHELLE.
CRFILT
radx,rady,mthresh.
CCDFILT
read-out noise (in e-), inverse gain factor (e-/ADU), threshold
(in units of standard deviation).
2|c| Parameters: RESPONSE command
Parameter
Description
STD
Raw Standard star, in STANDARD reduction mode.
RESPONSE
Computed instrument response in STANDARD reduction.
If the name NULL is assigned to this parameter no
response correction is applied in the REDUCE command.
Instead, the blaze function is fitted.
FLUXTAB
Table with absolute fluxes for the Standard star,
in STANDARD reduction mode. The tables are in the area
MID_STANDARD
FILTMED
Radius in x and y, threshold, as in command FILTER/MEDIAN
used to smooth the response function.
FILTSMO
Radius in x and y, threshold, as in command FILTER/SMOOTH
used to smooth the response function.
PIXNUL
Number of pixels to set to zero at the edges of each order.
BKGxxx
All background parameters : see definitions in Table 7.8)
FFOPT
Flat-Fielding option (YES/NO)
CORRECT
Flat field correction image (required if FFOPT=YES).
BLAZE
Smoothed flat field (required if FFOPT=YES).
EXTMTD
(See definition in Table 7.7)
SLIT
(See definition in Table 7.7)
OFFSET
(See definition in Table 7.7)
SAMPLE
(See definition in Table 7.7)
2|c| Parameters: REDUCE command
Parameter
Description
BKGxxx
All background parameters : see definitions in Table 7.8)
FFOPT
Flat-Fielding option (YES/NO)
CORRECT
Flat field correction image (required if FFOPT=YES).
BLAZE
Smoothed flat field (required if FFOPT=YES).
EXTMTD
(See definition in Table 7.7)
SLIT
(See definition in Table 7.7)
OFFSET
(See definition in Table 7.7)
SAMPLE
(See definition in Table 7.7)
RESPOPT
Option for the response corection (YES/NO).
RESPMTD
Method for the instrumental response correction (STD, IUE).
RESPONSE
Name of the response image as produced by RESPONSE/ECHELLE
(required if RESPMTD=STD).
RIPMTD
Defines the ripple correction algorithm as SINC for the method
fitting the blaze function, described by Ahmad(1981), and as
OVER for the method using the overlapping region of the orders,
proposed by Barker (1984) (required if RESPMTD=IUE).
RIPK
Defines the initial guess for the parameter k
in Equation 7.5.
The default value is updated during the wavelength calibration.
ALPHA
Defines the initial guess for the parameter in
Equation 7.5.
MGOPT
Merging option (YES/NO).
MRGMTD
Merging method (AVERAGE, NOAPPEND).
DELTA
Interval (in wavelength units) to be skipped at the edges of each order.