Health Check monitor


CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
AMBER
P2VM Low Res
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+R to enforce refresh of scores and dates]
AMBER trending system: HEALTH CHECK report
Last update: 2012-02-11T04:46:06 (UT)
| now: 2012-02-11T05:32:52 (UT)
HELP USERS-GUIDE MORE
same group: P2VM_LJHK K_P2VM_LJHK H_P2VM_LJHK J_P2VM_LJHK
General news:
AMBER news:
Report news: oct2011: the low threshold is now set to 95% of the median
Bas12: 0.765 to 1
Bas23: 0.754 to 1
Bas31: 0.676 to 1
DATE*: [?] 2012-02-04
report | NR
2012-02-05
report | NR
2012-02-06
report | NR
2012-02-07
report | NR
2012-02-08
report | NR
2012-02-09
report | NR
2012-02-10
report | NR
P...
Product quality:
products products products products products products products
  last OPSLOG data:
2012-02-10*

monitor

scores&comments | history ... | plot tutorial ... | contact                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
OPS?
**
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB yes (0) MEDIAN 0.82166 NONE VAL0.765,1 76 meas_int_Vis_base12 this | last_yr | all VIS12
2 QC1DB yes (0) MEDIAN 0.79892 NONE VAL0.754,1 76 meas_int_Vis_base23 this | last_yr | all VIS23
3 QC1DB yes (0) MEDIAN 0.72603 NONE VAL0.676,1 76 meas_int_Vis_base31 this | last_yr | all VIS31
*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: to indicate that OPSLOG data are included
Plot 1
scores: 6|
data source:amber_p2vm
(QC1 database)
dataset:meas_int_Vis_base12
median:0.82166NONE
fixed thresholds:0.765...1.0NONE
N_data plotted:76
[click on plot for closeup]
Plot 2
scores: 6|
data source:amber_p2vm
(QC1 database)
dataset:meas_int_Vis_base23
median:0.79892NONE
fixed thresholds:0.754...1.0NONE
N_data plotted:76
[click on plot for closeup]
Plot 3
scores: 6|
data source:amber_p2vm
(QC1 database)
dataset:meas_int_Vis_base31
median:0.72603NONE
fixed thresholds:0.676...1.0NONE
N_data plotted:76
[click on plot for closeup]
This plot

The P2VM (Pixel to Visibility Matrix) is used to calibrate the science data. The pipeline product provides also some QC parameters which are used to monitor the Instrumental transfer function and the offset between the photometric beams.
In Low resolution JHK, the Instrumental Visibility is calculated across the full JHK spectrum (this plot) or in the individual bands (J, H and K). The band K,H,J Instrumental Visibilities are shown in the HC plots: K_P2VM_LJHK, H_P2VM_LJHK, J_P2VM_LJH.
The HC plot "with_temp" is a long range plot where the temperature inside the lab is also plotted. This could show some correlation etween internal Visibility and temperature (see also environment HC plots under Non Daily)  


General information

Click on any of the plots to see a close-up version.

If applicable, the latest values from Paranal ops logs are plotted as blue dots. More information about those values, including a download option, is available in the OPS column.

The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up