|
|
 |
| |
| QC PLOTS |
| |
CURRENT |
HISTORY |
| dark current (until 2002-03) |
 |
 |
| dark current (since 2002-04) |
 |
 |
| QC1 database (advanced users):
browse
|
plot
|
Click on CURRENT to see the current (last) trending (Health Check).
Click on HISTORY to see the historical evolution of the trending. |
QC1 parameters
- dark current averaged over frame (e-/px/hr) (QC1 database table fors2_dark, column dark_avg)
This is the median level of the master dark, normalized per pixel and hour and converted into electrons.
- error of average dark current (e-/px/hr) (QC1 database table fors2_dark, column sigma_dark_avg)
As the error of the dark current is dominated by the read-out noise, it is calculated from the read-out noise provided in the header, scaled with the exposure time (as it is normalized to 1 hour).
History
For the old CCD (in use until March 22, 2002) the dark current was computed from the median master dark frame level in ADU/hr/px and transformed into electrons by using the conversion factor CONAD
(in the case of 4 port readout a mean CONAD was used). Also the dark current per port was calculated and stored in the QC1 Database (dark_A(BCD), sigma_dark_A(BCD)).
The high DARK values observed after January 2008 are due to pickup noise according to Thomas Szeifert (statement from May 3rd, 2008). |