mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
SINFONI:
lamp flat
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
SINFONI trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2013-05-25T05:01:18 (UT)
  | now: 2013-05-25T16:33:47 (UT) muc04 QC pipeline: sinfo-2.3.3 (installed 2013-04-15)
same group: lamp_eff   lamp_badpix lamp_off lamp_fpn1 lamp_fpn2 vignetting_check
General news: Click 'scores&comments' to switch to score details, and 'HC plot' to switch back to the graphics.
SINFONI news: PPRS050265: filter wheel position: DARKs are not valid (dome emission)
CCB-0000498/PPRS 049255 0500719: grating wheel failure: number of grating motions minimized
Report news:
DATE*: [?] 2013-05-18
report | NLT
2013-05-19
report | NLT
2013-05-20
report | NLT
2013-05-21
report | NLT
2013-05-22
report | NLT
2013-05-23
report | NLT
2013-05-24
report | NLT
P...
Product quality:
no products no products products products products products no products
  no OPSLOG data  

scores&comments | FULL | history ... | plot tutorial ... | contact               [?]
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
OPS?
**
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 oQC1DB no MEDIAN 1.42e+04 Number VAL2000,25000 8 qc_bpm_nbadpix this | last_yr | all number of bad pixels in the J-band flat at all camera scales
(this FULL data set is scored)
1 QC1DB no none   Number none  4 qc_bpm_nbadpix this | last_yr | all number of bad pixels: J-band/250 mas scale
1 QC1DB no none   Number none  0 qc_bpm_nbadpix this | last_yr | all number of bad pixels: J-band/100 mas camera scale
1 QC1DB no none   Number none  4 qc_bpm_nbadpix this | last_yr | all number of bad pixels: J-band/25 mas camera scale

2 oQC1DB no MEDIAN 1.68e+04 Number VAL2000,25000 24 qc_bpm_nbadpix this | last_yr | all number of bad pixels in the J-band flat at all camera scales
(this FULL data set is scored)
2 QC1DB no none   Number none  15 qc_bpm_nbadpix this | last_yr | all number of bad pixels: H-band/250 mas camera scale
2 QC1DB no none   Number none  9 qc_bpm_nbadpix this | last_yr | all number of bad pixels: H-band/100 mas camera scale
2 QC1DB no none   Number none  0 qc_bpm_nbadpix this | last_yr | all number of bad pixels: H-band/25 mas camera scale

3 oQC1DB no MEDIAN 1.26e+04 Number VAL2000,25000 109 qc_bpm_nbadpix this | last_yr | all number of bad pixels in the J-band flat at all camera scales
(this FULL data set is scored)
3 QC1DB no none   Number none  44 qc_bpm_nbadpix this | last_yr | all number of bad pixels: K-band/250 mas camera scale
3 QC1DB no none   Number none  44 qc_bpm_nbadpix this | last_yr | all number of bad pixels: K-band/100 mas camera scale
3 QC1DB no none   Number none  21 qc_bpm_nbadpix this | last_yr | all number of bad pixels: K-band/25 mas camera scale

4 oQC1DB no MEDIAN 1.02e+04 Number VAL2000,25000 51 qc_bpm_nbadpix this | last_yr | all number of bad pixels in the J-band flat at all camera scales
(this FULL data set is scored)
4 QC1DB no none   Number none  13 qc_bpm_nbadpix this | last_yr | all number of bad pixels: H+K-band/250 mas camera scale
4 QC1DB no none   Number none  19 qc_bpm_nbadpix this | last_yr | all number of bad pixels: H+K-band/100 mas camera scale
4 QC1DB no none   Number none  19 qc_bpm_nbadpix this | last_yr | all number of bad pixels: H+K-band/25 mas camera scale
*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: OPSLOG data are included
Plot 1
scores: 1|
data source:sinfoni_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_bpm_nbadpixo
median:1.42e+04Number
fixed thresholds:2000...25000Number
N_data plotted:8
[click on plot for closeup]
Plot 2
scores: 1|
data source:sinfoni_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_bpm_nbadpixo
median:1.68e+04Number
fixed thresholds:2000...25000Number
N_data plotted:24
[click on plot for closeup]
Plot 3
scores: 4|
data source:sinfoni_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_bpm_nbadpixo
median:1.26e+04Number
fixed thresholds:2000...25000Number
N_data plotted:109
[click on plot for closeup]
Plot 4
scores: 4|
data source:sinfoni_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_bpm_nbadpixo
median:1.02e+04Number
fixed thresholds:2000...25000Number
N_data plotted:51
[click on plot for closeup]
General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up