Health Check
monitor


Agora | CAL | HC | refs | QC
HealthCheck Monitor
    HOME | UsersGuide
SINFONI:
dark
Other HC:
QC links:
[best viewing with style sheets and javascript enabled]
SINFONI trending system: HEALTH CHECK plot
Last update: 2008-07-04T16:44:49 (UT)
now: 2008-07-06T06:07:38 (UT)
INFO USERS-GUIDE MORE
same group: DARK_dit=300   dit=10 dit=60 dit=120 dit=600 dit=900
last OPSLOG data:
2008-06-16
Plot: 1 2 3 4 5 6   latest_files
N_data: number of recent OPSLOG entries plotted here
N_data: 0 0 0 0 - - availability
history ... | plot tutorial ...   daily/often; important to check     [info]
Print: Information and research: News:
- png file - advanced studies: QC1 browser | QC1 plotter
 
Plot
?
Symb
?
Source
*
OPS?
**
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB yes (0) MEDIAN 10.33518 ADU 3SIG± 3.01808 50 qc_darkmed_ave this | last_yr | all median dark level, as computed from the master dark
2 QC1DB yes (0) MEDIAN 0.70727 ADU 3SIG± 1.1271 50 qc_darkmed_stdev this | last_yr | all standard deviation of median dark level
3 QC1DB yes (0) MEDIAN 3.02564 ADU 3SIG± 0.21505 50 qc_ron2 this | last_yr | all rms noise derived by subtracting 2 consecutive raw darks (normalized to DET.DIT = 1 second)
4 QC1DB yes (0) MEDIAN 2.37998 ADU 3SIG± 0.36048 50 qc_gaf_fpn this | last_yr | all fixed pattern noise as computed by fitting a Gaussian to the master dark histogram
5 QC1DB   MEDIAN 16631.0 number 3SIG± 2930.95923 50 qc_bpm_nbadpix this | last_yr | all number of pixels with a high dark current
6 QC1DB   MEDIAN -0.03923 unitless 3SIG± 0.94692 50 qc_oes2 this | last_yr | all median ratio of odd-even column stripes in the area of slitlet #25
*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: to indicate that OPSLOG data are included
This plot

These are the trending plots for the SINFONI QC parameters derived from DARK frames.

These plots describe:

1. the median DARK level, as computed from each master dark frame.

2. the standard deviation of the median DARK level.

3. the root-mean-squared noise, as derived by subtracting two consecutive raw dark frames, computing a median standard deviation in
selected sub-windows, computing a median standard deviation in selected sub-windows in the difference image, and then normalizing to DET.DIT=1 second.

4. the fixed pattern noise, as computed by fitting a Gaussian to the histogram of the master dark frames. The fpn is simply the standard deviation of Gaussian.

5. the number of pixels with a high dark current.

6. the odd-even column stripes in the area of slitlet #25. If med(odd) = median of all odd columns in the area of slitlet #25
and med(even) = median of all even columns in the area of slitlet #25 then, this plot is defined as:
oes25 = 2 * (med(odd) - med(even)) / (med(odd) + med(even))

Find more information about SINFONI DARK frames here.

Other QC parameters related to detector properties (like contamination, linearity, gain) are trended here.  


General information

Click on any of the plots to see a close-up version.

If applicable, the latest values from Paranal ops logs are plotted as blue dots. More information about those values, including a download option, is available in the OPSLOG section on top of the page.

The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up