mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
FULL reports
SINFONI:
wavelength
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
SINFONI trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2017-04-30T05:03:05 (UT)
  | now: 2017-04-30T09:02:20 (UT) muc04 QC pipeline: sinfo-2.9.9 (installed 2017-04-05)
same group: wavelength res_power   slit_pos#17 fwhm_lines N_lines lamp_mean lamp_max
General news:
SINFONI news:
Report news:
DATE*: [?] 2017-04-23
report | NLT
2017-04-24
report | NLT
2017-04-25
report | NLT
2017-04-26
report | NLT
2017-04-27
report | NLT
2017-04-28
report | NLT
2017-04-29
report | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw raw
P...
Product quality:
products products products products products products products
 
transfer   ngas [?]

scores&comments | FULL | history ... | plot tutorial ... | contact                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 2.01e+03 unitless 5%± 101 23 qc_spec_res_med this | last_yr | all median resolving power from arc lamp spectral lines (QC script) in the J-band (250 mas camera scale)
2 QC1DB MEDIAN 2.57e+03 unitless 5%± 129 22 qc_spec_res_med this | last_yr | all median resolving power from arc lamp spectral lines (QC script) in the H-band (250 mas camera scale)
3 QC1DB MEDIAN 4.74e+03 unitless 5%± 237 26 qc_spec_res_med this | last_yr | all median resolving power from arc lamp spectral lines (QC script) in the K-band (250 mas camera scale)
4 QC1DB MEDIAN 1.79e+03 unitless 5%± 89.4 34 qc_spec_res_med this | last_yr | all median resolving power from arc lamp spectral lines (QC script) in the H+K-band (250 mas camera scale)
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 1|
data source:sinfoni_wave
(QC1 database)
dataset:qc_spec_res_med
median:2.01e+03unitless
rel. thresholds: ±5%1913.6826201...2115.1228959unitless
N_data plotted:23
[click on plot for closeup]
Plot 2
scores: 2|
data source:sinfoni_wave
(QC1 database)
dataset:qc_spec_res_med
median:2.57e+03unitless
rel. thresholds: ±5%2442.88850765...2700.03466635unitless
N_data plotted:22
[click on plot for closeup]
Plot 3
scores: 6|
data source:sinfoni_wave
(QC1 database)
dataset:qc_spec_res_med
median:4.74e+03unitless
rel. thresholds: ±5%4504.5040343...4978.6623537unitless
N_data plotted:26
[click on plot for closeup]
Plot 4
scores: no current data
data source:sinfoni_wave
(QC1 database)
dataset:qc_spec_res_med
median:1.79e+03unitless
rel. thresholds: ±5%1697.65488727...1876.35540173unitless
N_data plotted:34
[click on plot for closeup]
This plot

green dots: the resolution is calculated using the QC script  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up