Quality Control parameters and trending

QUALITY CONTROL
  HOME
MIDI QC
Trending & QC1
QC1 and trending
Pipeline
Service Mode
Data Management
QC links:
[best viewing with style sheets and javascript enabled]
QC1 LINKS QC1 PLOTS
CURRENT
Transfer_Function
Detector
Optical elements
Reference Beams
Instrumental Transfer Function
RON | Linearity
Transmission | Wavelength Calibration
Reference beams: High_sens | SCI_PHOT | OPEN |

Find here detailed information about the MIDI QC1 parameters and select plots. ASCII data from the QC1 database will be available in the future.

GENERAL INFORMATION


Some MIDI calibration data are presently monitored to extract information about their quality and the instrument performance. The quality information is extracted into so-called QC parameters.

All measured QC1 parameters are stored in a database. Their evolution with time is evaluated by the trending process.

Up to now, only the evolution of the Instrumental Transfer function has been followed on two different time scales : a night or a full period. For P74, we are measuring some QC1 parameters from the technical calibrations : REFPIX (position of the reference pixels of the beams), DSPTRN (transmission of the dispersive element), DETRON (detector Read out noise), DETLIN (detector linearity).
 
INSTRUMENT PROPERTIES TRENDED


The following parameters are monitored and used for instrument trending. The plots are available on the MIDI HC page.

detector

alignment

dispersive elements (GRISM or PRISM)

Instrumental Transfer Function

  • More information under the specific TF page

 