HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
UVES:
lamps: stability
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
UVES trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-03-19T06:06:10 (UT)
  | now: 2024-03-19T09:59:53 (UT) muc02 QC pipeline: uves-6.1.11 (installed 2023-06-22)
same group: FF_Lamps ThAr_Lamp
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
UVES news:  
Report news:
DATE*: [?] 2024-03-12
report | NLT
2024-03-13
report | NLT
2024-03-14
report | NLT
2024-03-15
report | NLT
2024-03-16
report | NLT
2024-03-17
report | NLT
2024-03-18
report | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw no CAL raw files
P...
CAL product quality:
products products products products products products no CAL products
 
transfer   ngas [?]

scores&comments | FULL | history ... | plot tutorial ... | contact | HC_docu... | quality events...               [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 28.6 counts/sec 25%± 7.15 61 mean_master this | last_yr | all Stability of LAMP1 (flatfield deuterium lamp), BLUE CCD, 346nm, 1x1
2 QC1DB MEDIAN 180 counts/sec 25%± 44.9 62 mean_master this | last_yr | all same for LAMP3 (FF lamp), same setups
3 QC1DB MEDIAN 97.2 counts/sec VALVALUE=100,170 49 mean_master this | last_yr | all same for LAMP4 (FF lamp), 437nm, 1x1

5 QC1DB MEDIAN 998 counts/sec 25%± 250 123 mean_master this | last_yr | all same for LAMP5 (FF lamp), RED UPPER CCD, 580nm, 1x1
6 QC1DB MEDIAN 1.28e+03 counts/sec 25%± 321 123 mean_master this | last_yr | all same for LAMP5, RED LOWER CCD, 580nm, 1x1
7 QC1DB MEDIAN 392 counts/sec 25%± 98.1 51 mean_master this | last_yr | all same for LAMP6 (FF lamp), RED UPPER CCD, 860nm, 1x1
8 QC1DB MEDIAN 317 counts/sec 25%± 79.3 51 mean_master this | last_yr | all same for LAMP6, RED LOWER CCD, 860nm, 1x1
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 1|
data source:uves_flat
(QC1 database)
dataset:mean_master
median:28.6counts/sec
rel. thresholds: ±25%21.4425...35.7375counts/sec
N_data plotted:61
[click plot for closeup]
Plot 2
scores: 1|
data source:uves_flat
(QC1 database)
dataset:mean_master
median:180counts/sec
rel. thresholds: ±25%134.800125...224.666875counts/sec
N_data plotted:62
[click plot for closeup]
Plot 3
scores: 1|
data source:uves_flat
(QC1 database)
dataset:mean_master
median:97.2counts/sec
fixed thresholds:100...170counts/sec
N_data plotted:49
[click plot for closeup]
Plot 5
scores: 3|
data source:uves_flat
(QC1 database)
dataset:mean_master
median:998counts/sec
rel. thresholds: ±25%748.872...1248.12counts/sec
N_data plotted:123
[click plot for closeup]
Plot 6
scores: 3|
data source:uves_flat
(QC1 database)
dataset:mean_master
median:1.28e+03counts/sec
rel. thresholds: ±25%962.5695...1604.2825counts/sec
N_data plotted:123
[click plot for closeup]
Plot 7
scores: 2|
data source:uves_flat
(QC1 database)
dataset:mean_master
median:392counts/sec
rel. thresholds: ±25%294.294...490.49counts/sec
N_data plotted:51
[click plot for closeup]
Plot 8
scores: 2|
data source:uves_flat
(QC1 database)
dataset:mean_master
median:317counts/sec
rel. thresholds: ±25%237.756...396.26counts/sec
N_data plotted:51
[click plot for closeup]
This plot

This is the trending report for the UVES flat lamps. It monitors the deuterium lamp and the four standard flat-field lamps. It is based on data taken roughly once a week. Trends in this plot probably indicate aging of the lamp.
These plots describe:

  1. the mean_master QC parameter for LAMP 1, the Deuterium lamp used for BLUE FLATs with settings CD#1, 300-360nm
  2. the mean_master QC parameter for LAMP 3, used for BLUE FLATs with settings CD#1, 300-400nm
  3. the mean_master QC parameter for LAMP 4, used for BLUE FLATs with settings CD#2, 330-480nm
  4. the mean_master QC parameter for LAMP 5, REDU CCD, used for RED FLATs with settings CD#3, 450-650nm (which also includes MOS slit flats for the 520 and 580 settings)
  5. the same for the REDL CCD
  6. the mean_master QC parameter for LAMP 6, REDU CCD, used for RED FLATs with settings CD#3, 650-750nm and CD#4, 600-950nm (which also includes MOS slit flats for the 860 setting)
  7. the same for the REDL CCD.

Find more information about UVES Echelle FLAT frames here and UVES Echelle Orderdef frames here.  
General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up