mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
UVES:
wavelength calib 580
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
UVES trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-03-19T06:21:25 (UT)
  | now: 2024-03-19T11:36:14 (UT) muc02 QC pipeline: uves-6.1.11 (installed 2023-06-22)
same group: Res_U Res_L   Disp_U Disp_L   NLines_U NLines_L   WLmin_U WLmin_L   WLmax_U WLmax_L
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
UVES news:  
Report news:
DATE*: [?] 2024-03-12
report | NLT
2024-03-13
report | NLT
2024-03-14
report | NLT
2024-03-15
report | NLT
2024-03-16
report | NLT
2024-03-17
report | NLT
2024-03-18
report | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw raw
P...
CAL product quality:
products products products products products products products
 
transfer   ngas [?]

scores&comments | FULL | history ... | plot tutorial ... | contact | HC_docu... | quality events...                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEAN 51.7 none OFF± 2.5 81 resol_med this | last_yr | all Resolution of ThAr Lamp, from MOS WAVE data, 580.0nm, 1x1, Plate 1, 7+1FIB Mode, Fibre 1, RED UPPER CCD
2 QC1DB MEAN 52.4 none OFF± 2.5 82 resol_med this | last_yr | all same, for fibre 2
3 QC1DB MEAN 52.7 none OFF± 2.5 82 resol_med this | last_yr | all same, for fibre 3
4 QC1DB MEAN 52.4 none OFF± 2.5 82 resol_med this | last_yr | all same, for fibre 4

5 QC1DB MEAN 52.3 none OFF± 2.5 82 resol_med this | last_yr | all same, for fibre 5 This plot is scored
6 QC1DB MEAN 52.3 none OFF± 2.5 82 resol_med this | last_yr | all same, for fibre 6
7 QC1DB MEAN 52.5 none OFF± 2.5 82 resol_med this | last_yr | all same, for fibre 7
8 QC1DB MEAN 52.1 none OFF± 2.5 82 resol_med this | last_yr | all same, for fibre 8
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: not implemented
data source:uves_fib_wave
(QC1 database)
dataset:resol_med
mean:51.7none
rel. thresholds: 49.189164058...54.189164058none
N_data plotted:81
[click plot for closeup]
Plot 2
scores: not implemented
data source:uves_fib_wave
(QC1 database)
dataset:resol_med
mean:52.4none
rel. thresholds: 49.9025213463...54.9025213463none
N_data plotted:82
[click plot for closeup]
Plot 3
scores: not implemented
data source:uves_fib_wave
(QC1 database)
dataset:resol_med
mean:52.7none
rel. thresholds: 50.1955913683...55.1955913683none
N_data plotted:82
[click plot for closeup]
Plot 4
scores: not implemented
data source:uves_fib_wave_scoreQC
(QC1 database)
dataset:resol_med
mean:52.4none
rel. thresholds: 49.8577637646...54.8577637646none
N_data plotted:82
[click plot for closeup]
Plot 5
scores: 7|
data source:uves_fib_wave_fib5
(QC1 database)
dataset:resol_med
mean:52.3none
rel. thresholds: 49.8066351...54.8066351none
N_data plotted:82
[click plot for closeup]
Plot 6
scores: not implemented
data source:uves_fib_wave
(QC1 database)
dataset:resol_med
mean:52.3none
rel. thresholds: 49.7510710756...54.7510710756none
N_data plotted:82
[click plot for closeup]
Plot 7
scores: not implemented
data source:uves_fib_wave
(QC1 database)
dataset:resol_med
mean:52.5none
rel. thresholds: 49.9543986293...54.9543986293none
N_data plotted:82
[click plot for closeup]
Plot 8
scores: not implemented
data source:uves_fib_wave
(QC1 database)
dataset:resol_med
mean:52.1none
rel. thresholds: 49.6452989793...54.6452989793none
N_data plotted:82
[click plot for closeup]
This plot

This is the trending report for FLAMES/UVES QC parameters for wavelength calibration. It is based on data taken with the RED UPPER CCD in setting 580, 7+1FIB, with Plate 1. These data are taken daily as part of the Health Checks.

Note: only Fibre 5 is considered in the score for this report.
The individual plots describe:

  • Fibre 1: The average of the resolution values measured for each line included in the final selection
    Note that in contrast to fibres 2-8 which are illuminated by the robotic fibre projector fed by a Thorium-Argon lamp, fibre 1 is illuminated by the simultaneous calibration wavelength calibration lamp.
  • Fibres 2-8: The average of the resolution values measured for each line included in the final selection.

Note there is currently NO monitoring for fibre 9, used in the 8FIB modes.

User Manual specifications:

WL min_wav max_wav NMaxFib Gap [nm]

520 414 621 6 1

580 476 684 8 5

860 660 1060 8 10 Find more information about FLAMES/UVES wavelength solution monitoring here.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up