UVES trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-05-06T14:20:17 (UT)
  | now: 2024-05-06T23:29:58 (UT) muc02 QC pipeline: uves-6.1.11 (installed 2023-06-22)
same group: FF_Lamps ThAr_Lamp
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
UVES news:  
Report news:
DATE*: [?] 2024-04-29
report | NLT
2024-04-30
report | NLT
2024-05-01
report | NLT
2024-05-02
report | NLT
2024-05-03
report | NLT
2024-05-04
report | NLT
2024-05-05
report | NLT
2024-05-06
no data | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw raw no CAL raw files
P...
CAL product quality:
products products products products products products products no CAL products
 
transfer   ngas [?]

scores&comments | FULL | history ... | plot tutorial ... | contact | HC_docu... | quality events...                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 17.6 counts/sec VAL15.8,21.8 224 int_avg this | last_yr | all Stability of ThAr lamp from BLUE 437nm, 1x1, 0.4arcs data
2 QC1DB MEDIAN 18.1 counts/sec VAL15.26,21.26 212 int_avg this | last_yr | all Stability of ThAr lamp from REDU 580nm, 1x1, 0.3arcs data
3 QC1DB MEDIAN 15.1 counts/sec VAL12.77,18.77 216 int_avg this | last_yr | all Stability of ThAr lamp from REDL 580nm, 1x1, 0.3arcs data
4 QC1DB MEDIAN 27.4 counts/sec VAL4,80 39 int_avg this | last_yr | all Stability of ThAr lamp from BLUE 346nm, 1x1, 1.0arcs data. This is not scored
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 10|
data source:uves_wave
(QC1 database)
dataset:int_avg
median:17.6counts/sec
fixed thresholds:15.8...21.8counts/sec
N_data plotted:224
[click plot for closeup]
Plot 2
scores: 10|
data source:uves_wave
(QC1 database)
dataset:int_avg
median:18.1counts/sec
fixed thresholds:15.26...21.26counts/sec
N_data plotted:212
[click plot for closeup]
Plot 3
scores: 10|
data source:uves_wave
(QC1 database)
dataset:int_avg
median:15.1counts/sec
fixed thresholds:12.77...18.77counts/sec
N_data plotted:216
[click plot for closeup]
Plot 4
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:int_avg
median:27.4counts/sec
fixed thresholds:4...80counts/sec
N_data plotted:39
[click plot for closeup]
This plot

This is the trending report for the UVES ThAr lamp (LAMP2). It monitors the average intensity of the emission lines, normalized by EXPTIME. It is based on daily Health check data 437nm, 0.4arcs slit, and 580nm, 0.3 arcs slit.
These plots describe:

  1. the average emission line intensity int_avg for LAMP 2, the ThAr lamp, measured in the BLUE (437nm)
  2. the same as measured on the REDU CCD (580nm),
  3. the same as measured on the REDL CCD (580nm).

Find more information about UVES Echelle WAVE frames here.  
General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up