XSHOOTER trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2026-05-11T22:36:18 (UT)
  | now: 2026-05-12T01:42:06 (UT) muc02 QC pipeline: xshoo-3.6.5a (installed 2023-09-18)
same group: bias_AGC dark_AGC detmon_AGC
General news: NOTE: The Health Check and calChecker services have been moved to qcFlow. These pages are only maintained for comparison and will be removed in 2026.
XSHOOTER news:  
Report news:
DATE*: [?] 2026-05-05
no data | NLT
2026-05-06
no data | NLT
2026-05-07
no data | NLT
2026-05-08
no data | NLT
2026-05-09
no data | NLT
2026-05-10
no data | NLT
2026-05-11
no data | NLT
r...
Raw CAL displays:
no CAL raw files no CAL raw files no CAL raw files no CAL raw files no CAL raw files no CAL raw files no CAL raw files
P...
CAL product quality:
no CAL products no CAL products no CAL products no CAL products no CAL products no CAL products no CAL products
 
transfer   ngas [?]

scores&comments | history ... | plot tutorial ... | contact | HC_docu... | quality events...               [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 1.78e+03 ADU 20%± 355 88 median_master this | last_yr | all  
2 QC1DB MEDIAN 0.706 ADU 80%± 0.565 88 struct_x1 this | last_yr | all  
3 QC1DB MEDIAN 4.56 ADU 60%± 2.73 88 ron this | last_yr | all  
4 QC1DB MEDIAN 0.86 ADU 60%± 0.516 88 struct_y1 this | last_yr | all  
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 8|
data source:xshooter_bias
(QC1 database)
dataset:median_master
median:1.78e+03ADU
rel. thresholds: ±20%1420.9095704...2131.3643556ADU
N_data plotted:88
[click plot for closeup]
Plot 3
scores: 8|
data source:xshooter_bias
(QC1 database)
dataset:ron
median:4.56ADU
rel. thresholds: ±60%1.8232636...7.2930544ADU
N_data plotted:88
[click plot for closeup]
Plot 2
scores: not implemented
data source:xshooter_bias
(QC1 database)
dataset:struct_x1
median:0.706ADU
rel. thresholds: ±80%0.141262...1.271358ADU
N_data plotted:88
[click plot for closeup]
Plot 4
scores: not implemented
data source:xshooter_bias
(QC1 database)
dataset:struct_y1
median:0.86ADU
rel. thresholds: ±60%0.3440322...1.3761288ADU
N_data plotted:88
[click plot for closeup]
This plot

This plots shows detector related QC parameters derived from BIAS health check calibrations of the xshooter acquisition and guiding camera (AGC). It is a 562 x 528 pixel technical CCD. For the definition of the QC parameters, please see the corresponding section of the UVB-arm and VIS-arm science detector. On 2014-01-09 the detector operations mode was switched from DET.OUT1.GAININD = 8 to DET.OUT1.GAININD = 4.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up