mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
FORS2:
detector: monitoring
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
FORS2 trending system: FULL report
HELP USERS-GUIDE MORE
Last update: 2013-06-11T09:11:01 (UT)
  | now: 2025-09-10T16:33:31 (UT) muc01 QC pipeline: fors2-4.9.20 (installed 2013-04-15)
same group: dark linearity gain flux_range   conad fp._noise
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
FORS2 news:  
Report news: [an error occurred while processing this directive]
[an error occurred while processing this directive]
DATE*: [?] 2025-09-03
report | NLT
2025-09-04
report | NLT
2025-09-05
report | NLT
2025-09-06
report | NLT
2025-09-07
report | NLT
2025-09-08
report | NLT
2025-09-09
report | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw raw
P...
CAL product quality:
products products products products products products products
  no OPSLOG data  
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
OPS?
**
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB no none   e-/ADU none  61 gain all (full time range) gain for chip 1, high gain, R_SPECIAL
1 QC1DB no none   e-/ADU none  23 gain all (full time range) gain for chip 1, high gain, non-R_SPECIAL
1 oQC1DB no MEDIAN 0.715 e-/ADU VAL0.68,0.74 84 gain all (full time range) gain for chip 1, high gain, all filters
2 QC1DB no none   e-/ADU none  60 gain all (full time range) gain for chip 2, high gain, R_SPECIAL
2 QC1DB no none   e-/ADU none  23 gain all (full time range) gain for chip 2, high gain, non-R_SPECIAL
2 oQC1DB no MEDIAN 0.706 e-/ADU VAL0.68,0.74 83 gain all (full time range) gain for chip 2, high gain, all filters
*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: OPSLOG data are included
Plot 1
scores: [an error occurred while processing this directive]
data source:fors2_detmon
(QC1 database)
dataset:
(numbers below apply to this dataset)
gain
average:none
thresholds:none
N_data plotted:61
[click on plot for closeup]
Plot 2
scores: [an error occurred while processing this directive]
data source:fors2_detmon
(QC1 database)
dataset:
(numbers below apply to this dataset)
gain
average:none
thresholds:none
N_data plotted:60
[click on plot for closeup]
This plot

This is the DETMON GAIN health check plot. The gain values (units e-/ADU) are calculated by the detmon_opt_lg recipe (using the Photon Transfer Curve - PTC, method). The data come from a dedicated detmon calibrations consisting of sets of paired flat fields and bias frames with identical exposure time.

More information can be found on the plot tutorial page

TPL.ID: FORS2_img_tec_detLin
DB Name: fors2_detmon

 
General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up