mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
UVES:
wavelength calib daily CD2 CD3
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
UVES trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-04-26T11:24:44 (UT)
  | now: 2024-04-26T12:58:34 (UT) muc02 QC pipeline: uves-6.1.11 (installed 2023-06-22)
same group: Resolution   B_CD2_A B_CD2_B RU_CD3_A RU_CD3_B RL_CD3_A RL_CD3_B Thermal FWHM
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
UVES news:  
Report news:
DATE*: [?] 2024-04-19
report | NLT
2024-04-20
report | NLT
2024-04-21
report | NLT
2024-04-22
report | NLT
2024-04-23
report | NLT
2024-04-24
report | NLT
2024-04-25
report | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw raw
P...
CAL product quality:
products products products products products products products
 
transfer   ngas [?]

scores&comments | history ... | plot tutorial ... | contact | HC_docu... | quality events...                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 3.74e+03 Angstrom OFF± 10 85 lambda_min this | last_yr | all Minimum wavelength in A, from ARCLAMP data BLUE 437.0nm, 1x1, 0.4arcs slit
2 QC1DB MEDIAN 94 Angstrom OFF± 1 85 order_min_det this | last_yr | all Minimum order number, same data
3 QC1DB MEDIAN 0 Angstrom VAL-0.0001,0.0001 85 resid_avg this | last_yr | all Average residual in A, same data
4 QC1DB MEDIAN 4.99e+03 Angstrom OFF± 20 85 lambda_max this | last_yr | all Maximum wavelength in A, same data
5 QC1DB MEDIAN 124 Angstrom OFF± 1 85 order_max_det this | last_yr | all Maximum order number, same data
6 QC1DB MEDIAN 0.472 milli-Angstrom VAL0.4,3 85 resid_rms this | last_yr | all RMS of residuals in mA, same data
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 8|
data source:uves_wave
(QC1 database)
dataset:lambda_min
median:3.74e+03Angstrom
rel. thresholds: 3731.184...3751.184Angstrom
N_data plotted:85
[click plot for closeup]
Plot 2
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:order_min_det
median:94Angstrom
rel. thresholds: 93.0...95.0Angstrom
N_data plotted:85
[click plot for closeup]
Plot 3
scores: 8|
data source:uves_wave
(QC1 database)
dataset:resid_avg
median:0Angstrom
fixed thresholds:-0.0001...0.0001Angstrom
N_data plotted:85
[click plot for closeup]
Plot 4
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:lambda_max
median:4.99e+03Angstrom
rel. thresholds: 4973.748...5013.748Angstrom
N_data plotted:85
[click plot for closeup]
Plot 5
scores: 8|
data source:uves_wave
(QC1 database)
dataset:order_max_det
median:124Angstrom
rel. thresholds: 123.0...125.0Angstrom
N_data plotted:85
[click plot for closeup]
Plot 6
scores: 8|
data source:uves_wave
(QC1 database)
dataset:resid_rms
median:0.472milli-Angstrom
fixed thresholds:0.4...3milli-Angstrom
N_data plotted:85
[click plot for closeup]
This plot

This is the trending report for the quality of the UVES Echelle wavelength calibration. It is based on QC parameters derived from WAVE_ECH frames.

This report presents the first (summary A) of two sets of detailed parameters. Data are for the BLUE 437, 1x1, 0.4 arcs setting.

The individual plots describe:

  1. the minimum wavelength of the solution. Changes of a few nm usually occur because of changes in the last detected line, which often happens when the position of the orders changes slightly (see Grating Position trending). Changes of tens-hundreds of nm are usually due to the pipeline failing to locate entire orders.
  2. the minimum order found in the solution. (The minimum order is available in the QC1 database only since 2011-05-05.)
  3. The average of the residuals of the lines used in the final solution with respect to their laboratory wavelengths.
  4. the maximum wavelength of the solution (see also remark for the minimum wavelength)
  5. the maximum order found in the solution.
  6. the RMS of the residuals of the lines used in the final solution with respect to their laboratory wavelengths.

Find more information about UVES wavelength solution monitoring here.  
General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up