mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
MUSE:
detector:bias
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
MUSE trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-03-18T16:00:13 (UT)
  | now: 2024-03-19T03:20:45 (UT) muc09 QC pipeline: muse-2.8.8 (installed 2023-06-22)
same group: RON_master RON_det BIAS_raw_mean BIAS_raw_median BIAS_median temp
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
MUSE news:  
Report news:
DATE*: [?] 2024-03-12
report | NLT
2024-03-13
report | NLT
2024-03-14
report | NLT
2024-03-15
report | NLT
2024-03-16
report | NLT
2024-03-17
report | NLT
2024-03-18
no data | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw no CAL raw files
P...
CAL product quality:
products products products products products products no CAL products
 
transfer   ngas [?]

scores&comments | history ... | plot tutorial ... | contact | HC_docu... | quality events...                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB none   electrons VAL1.5,3.2 88 master1_ron this | last_yr | all average RON in quadrant 1
2 QC1DB none   electrons VAL1.5,3.2 88 master2_ron this | last_yr | all average RON in quadrant 2
3 QC1DB none   electrons VAL1.5,3.2 88 master3_ron this | last_yr | all average RON in quadrant 3
4 QC1DB none   electrons VAL1.5,3.2 88 master4_ron this | last_yr | all average RON in quadrant 4
5 QC1DB none   electrons VAL0.,0.8 88 master1_ron this | last_yr | all rms of the average RON in quadrant 1
6 QC1DB none   electrons VAL0.,0.8 88 master2_ron this | last_yr | all rms of the average RON in quadrant 2
7 QC1DB none   electrons VAL0.,0.8 88 master3_ron this | last_yr | all rms of the average RON in quadrant 3
8 QC1DB none   electrons VAL0.,0.8 88 master4_ron this | last_yr | all rms of the average RON in quadrant 4
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 7|
data source:muse_bias
(QC1 database)
dataset:master1_ron
average:none
fixed thresholds:1.5...3.2electrons
N_data plotted:88
[click plot for closeup]
Plot 2
scores: 7|
data source:muse_bias
(QC1 database)
dataset:master2_ron
average:none
fixed thresholds:1.5...3.2electrons
N_data plotted:88
[click plot for closeup]
Plot 3
scores: 7|
data source:muse_bias
(QC1 database)
dataset:master3_ron
average:none
fixed thresholds:1.5...3.2electrons
N_data plotted:88
[click plot for closeup]
Plot 4
scores: 7|
data source:muse_bias
(QC1 database)
dataset:master4_ron
average:none
fixed thresholds:1.5...3.2electrons
N_data plotted:88
[click plot for closeup]
Plot 5
scores: 7|
data source:muse_bias
(QC1 database)
dataset:master1_ron
average:none
fixed thresholds:0.0...0.8electrons
N_data plotted:88
[click plot for closeup]
Plot 6
scores: 7|
data source:muse_bias
(QC1 database)
dataset:master2_ron
average:none
fixed thresholds:0.0...0.8electrons
N_data plotted:88
[click plot for closeup]
Plot 7
scores: 7|
data source:muse_bias
(QC1 database)
dataset:master3_ron
average:none
fixed thresholds:0.0...0.8electrons
N_data plotted:88
[click plot for closeup]
Plot 8
scores: 7|
data source:muse_bias
(QC1 database)
dataset:master4_ron
average:none
fixed thresholds:0.0...0.8electrons
N_data plotted:88
[click plot for closeup]
This plot

This is MUSE RON_master health check plot. The read-out noise (RON) is calculated from the master BIAS calibrations, for each quadrant of each detector separately. Only BIAS calibrations taken with the read out mode (DET.READ.CURNAME) SCI1.0 are selected. The values averaged over all detectors for each quadrant are plotted in the upper panels. Also, the rms of the average values are plotted in the bottom panels.

TPL.ID: MUSE_cal_bias
Fits header keyword: QC.BIAS.MASTER(1/2/3/4).RON
DB Name: muse_bias
QC1 DB parameter: master(1/2/3/4)_ron

 
General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up