Health Check monitor


Agora | CAL | HC | refs | QC
HealthCheck Monitor
    HOME | UsersGuide
CRIRES:
wave cal
Other HC:
QC links:
[page auto-refreshes after 300 sec]            [press Ctrl+R to enforce refresh of scores and dates]
CRIRES trending system: SCORES (quick-look)
Last update: 2010-03-16T12:59:56 (UT)
| now: 2010-03-16T17:39:22 (UT)
HELP USERS-GUIDE MORE
same group: xcorr ThAr N2O DISP RESOL   AVG
last OPSLOG data:
2010-03-15*
Plot: 1 2 3 4 5 6 7 8   latest_files

N_data: number of recent OPSLOG entries plotted here
N_data: - - 6 - - - - - availability
HC plot... | history ... | plot tutorial ... | contact                 daily/often; important to check     [?] *Date on this monitor changes at 21:00 UT
WAVE dispersion (scores)
1: 1090.4_chip1
no current data
2: 1090.4_chip2
8|
3: 1090.4_chip3
8|
4: 1090.4_chip4
no current data
5: 4092.9_chip1
8|
6: 4092.9_chip2
8|
7: 4092.9_chip3
8|
8: 4092.9_chip4
8|
News:
 
Quality comments:
Plot Date* AB name score product quality comment
Plot 22010-03-12CRIRE.2010-03-13T13:20:22.860_tpl.ab
resolution found higher than expected
Plot 22010-03-14CRIRE.2010-03-15T10:33:15.727_tpl.ab
outlier dispersion DET3
Plot 32010-03-12CRIRE.2010-03-13T13:20:22.860_tpl.ab
resolution found higher than expected
Plot 32010-03-14CRIRE.2010-03-15T10:33:15.727_tpl.ab
outlier dispersion DET3
"Date" links to complete AB product page; "AB" links to score report


Scores (quick-look version)

This is the quick-look version of the HealthCheck Monitor. It displays scores for each HC report. Scores are flags assessing the compliance of data points with configured thresholds. A green score symbol indicates that the corresponding instrument component is performing as expected (OK) in the monitored time range.

If they exist, a table with quality comments (entered by the QC scientist) is displayed which usually (but not necessarily) refers to data with red scores. A comment may refer to the quality of the pipeline products and not necessarily to the exact parameter scored in this report. Quality comments are entered upon certification and, therefore, cannot be guaranteed to be available at the same time as the automatic scoring. These comments are also visible on the AB product monitors.

Scores are based on the QC data for the last 7 days before the indicated date (the last one with data for this report). For each HC plot, these scores are checked against the upper and lower thresholds, searching for outliers. If no outlier is found, the corresponding plot is scored OK and marked green. Outliers get a NOK score and are marked red. More ...