UVES trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-11-09T06:00:12 (UT)
  | now: 2024-11-09T17:45:25 (UT) muc02 QC pipeline: uves-6.1.11 (installed 2023-06-22)
same group: Median Master_RON Raw_RON Struct_X Struct_Y Pattern_noise
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
UVES news:  
Report news:
DATE*: [?] 2024-11-02
report | NLT
2024-11-03
report | NLT
2024-11-04
report | NLT
2024-11-05
report | NLT
2024-11-06
report | NLT
2024-11-07
report | NLT
2024-11-08
report | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw raw
P...
CAL product quality:
products products products products products products products
 
transfer   ngas [?]

scores&comments | FULL | history ... | plot tutorial ... | contact | HC_docu... | quality events...                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 192 ADU VAL27,325 88 median_master this | last_yr | all median level in MASTER BIAS, readout mode 225kHz,1x1,low; BLUE CCD
2 QC1DB MEDIAN 206 ADU VAL36,325 110 median_master this | last_yr | all same, for REDL CCD
3 QC1DB MEDIAN 209 ADU VAL62,325 110 median_master this | last_yr | all same, for REDU CCD
4 QC1DB MEDIAN 192 ADU VAL49,325 88 median_master this | last_yr | all median level in MASTER BIAS, readout mode 50kHz,2x2,high; BLUE CCD
5 QC1DB MEDIAN 208 ADU VAL33,325 95 median_master this | last_yr | all same, for REDL CCD
6 QC1DB MEDIAN 218 ADU VAL60,325 95 median_master this | last_yr | all same, for REDU CCD
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 7|
data source:uves_bias
(QC1 database)
dataset:median_master
median:192ADU
fixed thresholds:27...325ADU
N_data plotted:88
[click plot for closeup]
Plot 2
scores: 7|
data source:uves_bias
(QC1 database)
dataset:median_master
median:206ADU
fixed thresholds:36...325ADU
N_data plotted:110
[click plot for closeup]
Plot 3
scores: 7|
data source:uves_bias
(QC1 database)
dataset:median_master
median:209ADU
fixed thresholds:62...325ADU
N_data plotted:110
[click plot for closeup]
Plot 4
scores: 7|
data source:uves_bias
(QC1 database)
dataset:median_master
median:192ADU
fixed thresholds:49...325ADU
N_data plotted:88
[click plot for closeup]
Plot 5
scores: 8|
data source:uves_bias
(QC1 database)
dataset:median_master
median:208ADU
fixed thresholds:33...325ADU
N_data plotted:95
[click plot for closeup]
Plot 6
scores: 8|
data source:uves_bias
(QC1 database)
dataset:median_master
median:218ADU
fixed thresholds:60...325ADU
N_data plotted:95
[click plot for closeup]
This plot

This is the trending report for the BIAS level in UVES frames. Note that this parameter is based on the keyword PRO.DATAMED which is not present in the QC1 opslog files.
These plots describe:

  1. the BIAS level in masterbias frames of the BLUE CCD in the 1x1 binning, 225kHz, low gain readout mode
  2. same for RED LOWER CCD
  3. same for RED UPPER CCD
  4. the BIAS level in masterbias frames of the BLUE CCD in the 2x2 binning, 50kHz, high gain readout mode
  5. same for RED LOWER CCD
  6. same for RED UPPER CCD

The thresholds are set using the KPI document : "UVES critical Quality parameters"
The higher thrshold is always set to 325. The lower threshold depends of the Mode:
  1. B 1x1 : 27
  2. B 2x2 : 49
  3. U 1x1 : 33
  4. U 2x2 : 60
  5. L 1x1 : 36
  6. L 2x2 : 62
The reference values are the following
  1. B 1x1 : 195
  2. B 2x2 : 196
  3. U 1x1 : 204
  4. U 2x2 : 205
  5. L 1x1 : 202
  6. L 2x2 : 206
Find more information about UVES BIAS frames here.
Other QC parameters related to detector properties (like linearity, gain) are trended here.
 
General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up