UVES trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-04-24T18:15:01 (UT)
  | now: 2024-04-25T00:04:53 (UT) muc02 QC pipeline: uves-6.1.11 (installed 2023-06-22)
same group: Median Master_RON Raw_RON Struct_X Struct_Y Pattern_noise
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
UVES news:  
Report news:
DATE*: [?] 2024-04-17
report | NLT
2024-04-18
report | NLT
2024-04-19
report | NLT
2024-04-20
report | NLT
2024-04-21
report | NLT
2024-04-22
report | NLT
2024-04-23
report | NLT
2024-04-24
no data | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw raw no CAL raw files
P...
CAL product quality:
products products products products products products products no CAL products
 
transfer   ngas [?]

scores&comments | FULL | history ... | plot tutorial ... | contact | HC_docu... | quality events...                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 1.15 ADU VAL1.05,1.16 83 sigma_master this | last_yr | all RON in master BIAS frame, BLUE CCD, readout mode 225kHz,1x1,low br
2 QC1DB MEDIAN 1.63 ADU VAL1.58,1.74 124 sigma_master this | last_yr | all same, for REDL CCD
3 QC1DB MEDIAN 1.42 ADU VAL1.3,1.44 124 sigma_master this | last_yr | all same, for REDU CCD
4 QC1DB MEDIAN 2.22 ADU VAL2.09,2.32 80 sigma_master this | last_yr | all same as Fig. 1, for readout mode 50kHz,2x2,high
5 QC1DB MEDIAN 2.94 ADU VAL2.77,3.07 81 sigma_master this | last_yr | all same, for REDL CCD
6 QC1DB MEDIAN 2.52 ADU VAL2.49,2.76 81 sigma_master this | last_yr | all same, for REDU CCD
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 8|
data source:uves_bias
(QC1 database)
dataset:sigma_master
median:1.15ADU
fixed thresholds:1.05...1.16ADU
N_data plotted:83
[click plot for closeup]
Plot 2
scores: 14|
data source:uves_bias
(QC1 database)
dataset:sigma_master
median:1.63ADU
fixed thresholds:1.58...1.74ADU
N_data plotted:124
[click plot for closeup]
Plot 3
scores: 14|
data source:uves_bias
(QC1 database)
dataset:sigma_master
median:1.42ADU
fixed thresholds:1.3...1.44ADU
N_data plotted:124
[click plot for closeup]
Plot 4
scores: 7|
data source:uves_bias
(QC1 database)
dataset:sigma_master
median:2.22ADU
fixed thresholds:2.09...2.32ADU
N_data plotted:80
[click plot for closeup]
Plot 5
scores: 7|
data source:uves_bias
(QC1 database)
dataset:sigma_master
median:2.94ADU
fixed thresholds:2.77...3.07ADU
N_data plotted:81
[click plot for closeup]
Plot 6
scores: 7|
data source:uves_bias
(QC1 database)
dataset:sigma_master
median:2.52ADU
fixed thresholds:2.49...2.76ADU
N_data plotted:81
[click plot for closeup]
This plot

This is the trending report for the QC parameters derived from UVES BIAS frames.

These plots describe:

  • the read noise (RON) of the BLUE, RED LOWER and RED UPPER CCD master frame in the 1x1 binning, 225kHz, low gain readout mode;
  • the same for the 2x2 binning, 50kHz, high gain readout mode

The thresholds are set as -+5% of nominal values.

These nominal values are (in ADU)

  • BLUE 1x1 1.11
  • REDL 1x1 1.57
  • REDU 1x1 1.47
  • BLUE 2x2 2.21
  • REDL 2x2 2.93
  • REDU 2x2 2.63
The e- to ADU conversion used are
  • BLUE 1x1 1.84 ADU
  • REDL 1x1 1.59 ADU
  • REDU 1x1 1.46 ADU
  • BLUE 2x2 0.54 ADU
  • REDL 2x2 0.52 ADU
  • REDU 2x2 0.52 ADU
Find more information about UVES BIAS frames here.
Other QC parameters related to detector properties (like linearity, gain) are trended here.  
General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up