UVES trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-06-17T12:20:56 (UT)
  | now: 2024-06-17T12:30:04 (UT) muc02 QC pipeline: uves-6.1.11 (installed 2023-06-22)
same group: Resolution   B_CD2_A B_CD2_B RU_CD3_A RU_CD3_B RL_CD3_A RL_CD3_B Thermal FWHM
close window history: 2007 2008 2009 2010 2011 2012 2013 2014 2015 2016 2017 2018 2019 2020 2021 2022 2023   2024   HEALTH
  ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 102 R VAL95,125 185 resol_med this | last_yr | all Average resolving power, from ARCLAMP data for REDU 580.0nm, 1x1, 0.3arcs slit. Lower threshold normally set to the value in the user manual
2 QC1DB MEDIAN 18.5 ADU/s VAL15.26,21.26 176 int_avg this | last_yr | all Average intensity of selected lines, same data
3 QC1DB MEDIAN 1.11e+03 Number none  98 nlin_tot this | last_yr | all Number of lines found, same data
4 QC1DB MEDIAN 10.2 R VAL0,14 178 resol_rms this | last_yr | all RMS of resolving power, same data
5 QC1DB FIXED 34 Number none  98 nlin_int this | last_yr | all Number of lines used in Integrated Average Intensity, same data
6 QC1DB MEDIAN 490 Number VAL310,550 98 nlin_sol this | last_yr | all Number of lines actually used in final solution, same data
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 8|
data source:uves_wave
(QC1 database)
dataset:resol_med
median:102R
fixed thresholds:95...125R
N_data plotted:185
[click plot for closeup]
Plot 2
scores: 8|
data source:uves_wave
(QC1 database)
dataset:int_avg
median:18.5ADU/s
fixed thresholds:15.26...21.26ADU/s
N_data plotted:176
[click plot for closeup]
Plot 3
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:nlin_tot
median:1.11e+03Number
thresholds:none 
N_data plotted:98
[click plot for closeup]
Plot 4
scores: 8|
data source:uves_wave
(QC1 database)
dataset:resol_rms
median:10.2R
fixed thresholds:0...14R
N_data plotted:178
[click plot for closeup]
Plot 5
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:nlin_int
fixed average:34Number
thresholds:none 
N_data plotted:98
[click plot for closeup]
Plot 6
scores: 8|
data source:uves_wave
(QC1 database)
dataset:nlin_sol
median:490Number
fixed thresholds:310...550Number
N_data plotted:98
[click plot for closeup]
This plot

This is the trending report for monitoring UVES Resolution based on QC parameters derived from Wavelength Calibration frames.
This report presents a second set of detailed parameters of the wavelength solution for the UPPER chip of the RED 580.0nm, 1x1 binning data, 4pts/625kHz/lg readout mode, 0.3 arcsec slit-width setup.
The individual plots describe:

  1. The average of the resolution values measured for each line included in the final selection, (this is a repeat of the data presented in the RESOLUTION overview report. For this plot the value of the lower threshold is set to 95% of the value in the user manual because of charge diffusion.
  2. Average intensity of a subset of lines scaled to unit exposure time
  3. The total number of lines found in the frame. The PL automatically adjusts the detection threshold until the number of lines found falls within the --minlines and --maxlines parameters of the recipe, by default 1000 and 1400 for the REDU chip
  4. The RMS of the resolution values measured for each line included in the final selection
  5. The number of lines included in the calculation of the average of the intensity
  6. The number of lines included in the final wavelength calibration solution. The selection is made in a two step process, first line with residuals larger than --tolerance (default 0.6 pix) are discarded, then the remaining list is further refined with kappa-sigma clipping controlled by the --kappa (default 4.0 sigma) recipe command line option.

SLITW R_blue R_red(L) old_R_red(U)
#2.1 20000 20000 19000
#2.0 22000 22000 20900
#1.50 31000 31000 29450
#1.40 32000 32000 30400
#1.30 34000 34000 32300
#1.20 37000 37000 35150
#1.10 39000 39000 37050
#1.00 41000 42000 39900
#0.90 44000 46000 43700
#0.80 50000 53000 50350
#0.70 54000 60000 57000
#0.60 59000 68000 64600
#0.50 65000 78000 74100
#0.40 71000 91000 86450
#0.30 78000 108000 102600
#0.20 82000 129000 122550
Find more information about UVES Wavelength Solution monitoring here.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up