mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
ISAAC:
detector: monitoring
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
ISAAC trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2013-12-16T09:56:34 (UT)
  | now: 2024-05-02T21:48:21 (UT) muc03 QC pipeline: isaac-6.1.3 (installed 2013-04-15)
same group: SW_FPN LW_DH_FPN LW_DL_FPN LW_UH_FPN SW_STD LW_DH_STD LW_DL_STD LW_UH_STD
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
ISAAC news: 2013-12-16: due to ISAAC decommissioning, refreshing of the health checks disabled
Report news:
DATE*: [?] 2013-12-09
report | NLT
2013-12-10
report | NLT
2013-12-11
report | NLT
2013-12-12
report | NLT
2013-12-13
no data | NLT
2013-12-14
no data | NLT
2013-12-15
no data | NLT
P...
Product quality:
products products products products no products no products no products
  no OPSLOG data  

scores&comments | history ... | plot tutorial ... | contact                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
OPS?
**
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB no none   ADU none  49 qc_UL_stdev this | last_yr | all structural (fixed pattern) noise for DoubleCorrLowBias DIT=2.0 in upper left (UL) quadrant.
2 QC1DB no none   ADU none  49 qc_UR_stdev this | last_yr | all like Fig.1, upper right (UR) quadrant, DIT=2.0
3 QC1DB no none   ADU none  48 qc_UL_stdev this | last_yr | all structural (fixed pattern) noise for DoubleCorrLowBias DIT=5.0 in upper left (UL) quadrant.
4 QC1DB no none   ADU none  48 qc_UR_stdev this | last_yr | all like Fig.1, upper right (UR) quadrant, DIT=5.0
5 QC1DB no none   ADU none  49 qc_LL_stdev this | last_yr | all like Fig.1, lower left (LL) quadrant, DIT=2.0
6 QC1DB no none   ADU none  49 qc_LR_stdev this | last_yr | all like Fig.1, lower right (LR) quadrant, DIT=2.0
7 QC1DB no none   ADU none  48 qc_LL_stdev this | last_yr | all like Fig.1, lower left (LL) quadrant, DIT=5.0
8 QC1DB no none   ADU none  48 qc_LR_stdev this | last_yr | all like Fig.1, lower right (LR) quadrant, DIT=5.0
*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: OPSLOG data are included
Plot 1
scores: not implemented
data source:isaac_dark
(QC1 database)
dataset:qc_UL_stdev
average:none
thresholds:none
N_data plotted:49
[click on plot for closeup]
Plot 2
scores: not implemented
data source:isaac_dark
(QC1 database)
dataset:qc_UR_stdev
average:none
thresholds:none
N_data plotted:49
[click on plot for closeup]
Plot 3
scores: not implemented
data source:isaac_dark
(QC1 database)
dataset:qc_UL_stdev
average:none
thresholds:none
N_data plotted:48
[click on plot for closeup]
Plot 4
scores: not implemented
data source:isaac_dark
(QC1 database)
dataset:qc_UR_stdev
average:none
thresholds:none
N_data plotted:48
[click on plot for closeup]
Plot 5
scores: not implemented
data source:isaac_dark
(QC1 database)
dataset:qc_LL_stdev
average:none
thresholds:none
N_data plotted:49
[click on plot for closeup]
Plot 6
scores: not implemented
data source:isaac_dark
(QC1 database)
dataset:qc_LR_stdev
average:none
thresholds:none
N_data plotted:49
[click on plot for closeup]
Plot 7
scores: not implemented
data source:isaac_dark
(QC1 database)
dataset:qc_LL_stdev
average:none
thresholds:none
N_data plotted:48
[click on plot for closeup]
Plot 8
scores: not implemented
data source:isaac_dark
(QC1 database)
dataset:qc_LR_stdev
average:none
thresholds:none
N_data plotted:48
[click on plot for closeup]
This plot

This is the trending plot for the ISAAC Aladdin structural noise (fixed pattern noise) QC parameters derived from DARK frames.

QC parameter are derived from post-pipeline Quality Control processes in Garching, hence there are no values from Paranal to be included.

Method: take a simple standard deviation of the MASTER_DARK frame quadrant, no clipping is applied. This paramater is by intention sensitive to changes in the number and strength of hot pixels. The robust scatter of the MASTER_DARKs is trended by the FPN QC parameter

These plots describe:
1. stdev mode = DoubleCorrLowBias, DIT=2.0sec, quadrant = upper left (UL),
2. stdev As Fig.1, but for UR quadrant
5. stdev As Fig.1 but for LL quadrant
6. stdev As Fig.1 but for LR quadrant

3. stdev mode = DoubleCorrLowBias, DIT=5.0sec, quadrant = upper left (UL),
4. stdev As Fig.1, but for UR quadrant
7. stdev As Fig.1 but for LL quadrant
8. stdev As Fig.1 but for LR quadrant

Find more information about ISAAC DARK frames here.

Other QC parameters related to detector properties (like read noise, linearity) are trended.  


General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up