Evolution of the Response Function
The Response Function (RF) is computed during (photometric) technical or calibration nights. For this purpose spectrophotometric standars are observed with Grism 11 and a slit of 5". Because EFOSC2 uses an UV-flooded CCD, it is important to monitor the response in the blue region of the spectrum.The figure below shows RFs computed in different dates.
Notes:
- February 2010: standars where observed with the two off-centered slits, so the blue and red curves show the gain in spectral coverage that can be obtained with this method.
- First half of 2010: the ca. 24% lower response is due to dust accumulating on the main mirror of the NTT. After mirror washing the system efficiency has been restored (see 2010-dec-21).
- February 2012: dust is again progressively reducing the system efficiency, as ZPs in QC pages confirm.
-
General: For a finer time sampling of the efficiency
evolution, the zero-points computed by the pipeline can be inspected at the following link:
Quality Control .