| NAME | run_amber_processQC.csh | |||||||||||||||
| VERSION | 1.0 | |||||||||||||||
| INSTRUMENT | AMBER | |||||||||||||||
| SYNTAX | SHELL | |||||||||||||||
| RAWTYPE |
CALIB_FRINGE SCIENCE_FRINGE BPOS 3P2V | |||||||||||||||
| CALL |
processQC -a abname or . ~/Isabelle/SCRIPTS/MeasureQuality/run_amber_processQC.sh.averaged $RAW_TYPE $DATE $PF | |||||||||||||||
| PURPOSE | extracting various QC parameters | |||||||||||||||
| PROCINPUT | products under CALIB_FRINGE/SCIENCE_FRINGE/BPOS/3P2V | |||||||||||||||
| QC1TABLE |
trending | table(s) in QC1 database: amber_science_fringe > amber_bpos > amber_p2vm > amber_calibrators_K > amber_cutoff > amber_science_fringe > | |||||||||||||||
| TRENDPLOT |
trending | current trending plot(s) associated to this procedure: >trend_report_BPOS_shiftK_noBCD_HC.html | >trend_report_P2VM_LJHK_HC.html | >trend_report_WINDOW_SM_HC.html | >trend_report_SNR_HC.html | |||||||||||||||
| QC1PAGE |
trending | associated trending page(s): none | |||||||||||||||
| QC1PLOTS |
trend_report_BPOS_shiftK_noBCD_HC.png trend_report_P2VM_LJHK_HC.png trend_report_WINDOW_SM_HC.png trend_report_SNR_HC.png product_name.fits.gif | |||||||||||||||
| QC1PARAM |
QC1 parameters written into QC1 table: QC1db names: amber_bpos | amber_p2vm | amber_cutoff | amber_calibrators_K(JH) | amber_science_fringe (KHJ) FITS key names for BPOS: QC.REFPIX.A/B/C.K/H/J.SHX/SHY | QC.REFPIX.A/B/C.K/H/J.X/Y | QC.REFPIX.A/B/C.K/H/J.FLUX/PEAKFLUX/SIZX/SIZY/ANGLE FITS key names for P2VM: QC.P1/2/3.OFFSETX/Y | QC.P2VM.K/H/J.VIS12/VIS23/VIS31 | QC.P2VM.K/H/J.ERRVIS12/ERRVIS23/ERRVIS31 FITS key names for cutoff: CHANMAX/MIN | LAMBDAMAX/MIN | FITS key names for SNR (amber_calibrators_K/H/J and amber_science_fringeKHJ): QC.FRAMES.SELECTED.SNRGT2 | QC.FRAMES.SELECTED.PERCENT QC1 params created by pipeline: all (except cutoff) QC1 params created by this procedure: CHANMAX/MIN | LAMBDAMAX/MIN | |||||||||||||||
| ALGORITHM |
Description of algorithms: BPOS: HC calibration: under Position Ref. measure the XY position of the beam for the 3 beams ABC and the JHK bands. The shift between these different settings and the amber reference (coldstop) is calculated. BPOS: HC calibration: under Flux Angle Ref. measure the flux, the flux on the peak, the position angle and the size in X and Y of the beam for the 3 beams ABC and the JHK bands. BPOS: HC calibration: under LAMP. monitoring of the LAMP in (XY, FLUX and SIZZE) for beam A/B/C in J/H/K Description of algorithms: P2VM: Calculate the internal instrumental Visibility for the different instrumental settings (Low, Medium and High resolution) and for the 3 baselines (12, 23, 31). The Spectral Configuration is given by OCS.OBS.SPECCONF or INS.MODE, teh observing band by QC.BAND. When observing in Low Resolution JHK or Medium Resolution the offsets between the different photometric channels is also measured Description of algorithms: Cutoff: When fringes are obtained in Low Resolution, we monitor the cutoff of the K spectruum, this has been an issue with AMBER and needs to be monitor to insure that the full K spectruum is present in the raw data. Description of algorithms: SNR: When fringes are obtained, teh number of frames with a SNR >= 2 is calculated. | |||||||||||||||
| CERTIF |
(guidelines for acceptance/rejection): for the BPOS, P2VM, check if it is within the same range as the HC plots (guidelines for acceptance/rejection): for the fringe files, check the Vis2=f(SNR), the shape of the spectruum and the shape of the dispersed fringes (!! still fringes for MR and HR) | |||||||||||||||
| COMMENTS |
This run_amber_processQC.sh.averaged includes all the QC procedures To see the individual qcDocu comments, please check the individual documentations | |||||||||||||||