| This documentation is intended both for QC scientists and SciOps astronomers (who may want to ignore the technical information displayed in grey).
|
| |
|
|
|
|
| |
|
bottom
plots
|
contact
|
|
| NAME |
run_detron.csh
|
|
| VERSION |
1.0
|
|
| SYNTAX |
SHELL
|
|
| CALL |
processQC -a abname or
. ~/Isabelle/SCRIPTS/MeasureQuality/run_data_quality_processQC.csh $RAW_TYPE $DATE $PF
. ~/Isabelle/SCRIPTS/MeasureQuality/run_detron.csh
|
|
| INSTRUMENT |
MIDI
|
|
| RAWTYPE |
CAL_DETRON
|
|
| PURPOSE |
extracts values of the RON
|
|
| PROCINPUT |
products under CAL_DETRON
|
|
| QC1TABLE |
trending | table(s) in QC1 database:
midi_detron
|
|
| TRENDPLOT |
trending | HealthCheck plot(s) associated to this procedure:
trend_report_DETRON_HC.html
|
|
| QC1PAGE |
trending | associated documentation:
MIDI_QC_tutorial_detector.html
|
|
QC1PLOTS
top |
trend_report_DETRON_HC.png
shows the median level of the RON
|
|
| QC1PARAM |
QC1 parameters written into QC1 table:
QC1db names: midi_detron
FITS key names: QC.DETRON.MEDIAN
QC1 params created by pipeline: all
QC1 params created by this procedure: none
|
|
| ALGORITHM |
Description of algorithms:
QC.DETRON.MEDIAN: median of the standard deviation over all the pixels and all the frames (in ADU)
|
|
| CERTIF |
(guidelines for acceptance/rejection) should be within the same range than the HC plot
since last detector intervention in 2009: between 10-12
|
|
| COMMENTS |
any increase in the RON should be notified to Paranal (could be due to a power cut and condensation on the detector)
|
| top |
[2010-01-26T08:57:00] created by qcDocu v1.1.2, a tqs tool
|