NAME run_detron.csh
VERSION 1.0
SYNTAX SHELL
CALL processQC -a abname or
. ~/Isabelle/SCRIPTS/MeasureQuality/run_data_quality_processQC.csh $RAW_TYPE $DATE $PF
. ~/Isabelle/SCRIPTS/MeasureQuality/run_detron.csh
INSTRUMENT MIDI
RAWTYPE CAL_DETRON
PURPOSE extracts values of the RON
PROCINPUT products under CAL_DETRON
QC1TABLE trending | table(s) in QC1 database:
midi_detron
TRENDPLOT trending | HealthCheck plot(s) associated to this procedure:
trend_report_DETRON_HC.html
QC1PAGE trending | associated documentation:
MIDI_QC_tutorial_detector.html
QC1PLOTS
top

trend_report_DETRON_HC.png
shows the median level of the RON
QC1PARAM QC1 parameters written into QC1 table:
QC1db names: midi_detron
FITS key names: QC.DETRON.MEDIAN
QC1 params created by pipeline: all
QC1 params created by this procedure: none
ALGORITHM Description of algorithms:
QC.DETRON.MEDIAN: median of the standard deviation over all the pixels and all the frames (in ADU)
CERTIF (guidelines for acceptance/rejection) should be within the same range than the HC plot
since last detector intervention in 2009: between 10-12
COMMENTS any increase in the RON should be notified to Paranal (could be due to a power cut and condensation on the detector)