| QC PLOTS |
| |
CURRENT |
HISTORY |
| Daily Health Checks: |
|
|
| ECH FF lamp stability |
 |
 |
| ECH ThAr lamp stability |
 |
 |
| QC1 uves_flat database (advanced users):
browse |
plot |
|
|
Click on CURRENT to see the last trending (Health Check).
Click on HISTORY to see the historical evolution of the trending. |
The monitoring of the ECH calibration lamps is based on FLAT and WAVE calibration data mostly
fed by dedicated HC data.
All five FLAT lamps are monitored, as well as the ARC lamp LAMP2 which is based on daily Health
check data 437nm, 0.4arcs slit, and 580nm, 0.3 arcs slit.
mean_master: mean value across master
frame (QC1
database table uves_flat, column mean_master, product keyword PRO.DATAAVG)
Mean of the pixel values in the frame. All pixels are
taken into account.
int_avg: average intensity of the emission lines, normalized
by EXPTIME (QC1 database table uves_wave)
|