Testbench Operational Range, characterization and setup

MONOCHROMATOR 1 and 2
Bandwidth range is 0.1 to 20 nm (ORIEL 77722)


Usable range is 250-1600nm (GRATING ORIEL 77744)
The Sphere
This is a labsphere CSTM-US-200-SF :
The US-200-SF is a 20 inch diameter integrating sphere that provides exceptional uniform luminance over the full 8 inch diameter of the sphere exit port. The sphere interior is coated with Spectraflect coating. Four internal lamp mounts are equally spaced around the exit port aperture. The lamp mounts are designed to accept two 45 watt and two 150 watt direct mount halogen light sources. An additional 0.50 port is provided adjacent to the exit port to accommodate a luminance monitor. A plug is included if the port is not used. The mounting frame supplied with the US-200-SF allows for sphere height adjustment while allowing sphere rotation to position the exit port aperture in a vertical or horizontal plane

Neutral Filters
Till the 13/03/2002 : 100%, 50%, 25%, 10% and 1%
This is metal coated filters, 25mm, from Melles Griot
Order Filters
Oriel filters 51290 and 51330 are used to sort the orders:
- from 250 - 500nm: filter 1 (no filter)
- from 500 - 700nm: filter 2 (51290/cut-on wl: 475nm GG 475 SCHOTT cuts everythings below 475nm) , get here transmittance
- from 700 - 1100nm: filter 3 (51330/cut-on wl: 665nm RG 665 SCHOTT cuts everythings below 665nm), get here transmittance
Calibration curve using a ZINC LAMP


Close up around 330nm, very good agreement with ZINC spectral lamp.
LAMP
XENON
HALOGEN
SPHERE
If SPECTRALON is used, reflectivity is above 95% between 300 and 1.1um, 200nm , 90%
DIODE


WINDOWS


Overall photon output (where CCD is sitting) photons/cm2/s.

Ratio between 120 and 100 W, Overall output