Instrument's Characteristics
In this page are listed the main characteristic of XSHOOTER. However, this same page is intended as a quick reference only.
The instrument reference guide and complete document is - and will always be - the XSHOOTER User Manual. Do not forget to read it! The new resolving power values presented below for the UVB/VIS arms still need to be refined, in particular for the 1.0" UVB and 0.9" VIS slits.
spectrograph characteristics
| arm | λ-range (nm) | N. of orders | scale (''/pix) [1] | AB limit mag [2] |
|---|---|---|---|---|
| UVB | 300-560 | 12 | 0.16 | 21.2 (at 356.1 nm, ord N.21) 21.7 (at 438.8 nm, ord N.17) |
| VIS | 550-1020 | 15 | 0.16 | 20.9 (at 653.8 nm, ord N.35) 20.8 (at 777.6 nm, ord N.21) |
| NIR [3] | 1020-2480 | 16 | 0.21 | 21.0 (at 1245.2 nm, ord N.21) 20.6 (at 1634.4 nm, ord N.16) 18.7 (at 2179.2 nm, ord N.12) |
[2] The data collected during the UVB and VIS arm commissioning are not sufficient to provide reliable limit magnitudes. The limit magnitudes in the table are predicted by the XSHOOTER ETC which relies on the observed instrument efficiency. The accuracy of the ETC is currently limited to ~20% in the UVB and VIS arm, but is subject to continuous update and improvements. The posted values have been computed for a SNR ~ 10 in 1 hr integration time. Input parameters were: slit=1" (for the UVB, 0.9" for the VIS and the NIR), airmass= 1.2, seeing=0.8", 3 days from new moon and a power low spectrum with index =-2.
[3] NIR arm values are predicted by the instrument physical model.
New spectrograph resolution, especially for the NIR arm with the new slits wheel [1]
| arm | slit width ('') [2] | R=(Δλ/&lambda) | sampling (pix/FWHM) | arm | slit width ('') | R=(Δλ/&lambda) | sampling (pix/FWHM) | arm | slit width ('') | R=(Δλ/&lambda) | sampling (pix/FWHM) |
|---|---|---|---|---|---|---|---|---|---|---|---|
| UVB | 0.5 | 9900 | 3.2 | VIS | 0.4 | 18200 | 2.9 | NIR | 0.4 | 10500 | 2.2 |
| 0.8 | 6200 | 5.2 | 0.7 | 10600 | 4.9 | 0.6 | 7780 | 2.9 | |||
| 0.6JH [4] | 7760 | ||||||||||
| 1.0 | 4350 | 5.4 | 0.9 | 7450 | 7.1 | 0.9 | 5300 | 4.2 | |||
| 0.9JH [4] | 5300 | ||||||||||
| 1.3 | 4000 | 8.1 | 1.2 | 6700 | 7.9 | 1.2 | 3890 | 5.8 | |||
| 1.6 | 3300 | 9.9 | 1.5 | 5400 | 9.7 | ||||||
| UVB | IFU [3] | 8400 | 3.9 | VIS | IFU [3] | 13200 | 4.0 | NIR | IFU [3] | 8300 | 2.7 |
[2] All the slits have a fixed length of 11''.
[3] The IFU FoV is 4'' x 1.8 '' which are re-sampled in a 0.6'' x 12 '' slit in the spectrographs. [4] The 0.6"JH and 0.9"JH are the 2 new slits with the K band blocking filter
Old spectrograph resolution [1]
| arm | slit width ('') [2] | R=(Δλ/&lambda) | sampling (pix/FWHM) | arm | slit width ('') | R=(Δλ/&lambda) | sampling (pix/FWHM) | arm | slit width ('') | R=(Δλ/&lambda) | sampling (pix/FWHM) |
|---|---|---|---|---|---|---|---|---|---|---|---|
| UVB | 0.5 | 9100 | 3.5 | VIS | 0.4 | 17400 | 3.0 | NIR | 0.4 | 9100 | 3.5 |
| 0.8 | 6200 | 5.2 | 0.7 | 11000 | 4.8 | 0.6 | 6200 | 5.2 | |||
| 1.0 | 5100 | 6.3 | 0.9 | 8800 | 6.0 | 0.9 | 5100 | 6.3 | |||
| 1.3 | 4000 | 8.1 | 1.2 | 6700 | 7.9 | 1.2 | 4000 | 8.1 | |||
| 1.6 | 3300 | 9.9 | 1.5 | 5400 | 9.7 | 1.5 | 3300 | 9.9 | |||
| UVB | IFU [3] | 7900 | 4.1 | VIS | IFU [3] | 12600 | 4.2 | NIR | IFU [3] | 8100 | 2.6 |
[2] All the slits have a fixed length of 11''.
[3] The IFU FoV is 4'' x 1.8 '' which are re-sampled in a 0.6'' x 12 '' slit in the spectrographs.
detector characteristics
| arm | detector | readout mode | gain (e/ADU) | RON (e) | readout time (s) | saturation (ADU) |
|---|---|---|---|---|---|---|
| UVB | 4k x 2k E2V CCD44-82 | 100,1x1,hg | 0.67 | 2.6 | 70 | 65000 |
| 100,1x2,hg | 0.67 | 2.6 | 38 | |||
| 100,2x2,hg | 0.67 | 2.6 | 22 | |||
| 400,1x1,lg | 1.75 | 4.5 | 19 | |||
| 400,1x2,lg | 1.75 | 4.5 | 12 | |||
| 400,1x2,lg | 1.75 | 4.5 | 8 | |||
| VIS | 4k x 2k MIT/LL CCID 20 | 100,1x1,hg | 0.64 | 3.2 | 92 | 65000 |
| 100,1x2,hg | 0.64 | 3.2 | 48 | |||
| 100,2x2,hg | 0.64 | 3.2 | 27 | |||
| 400,1x1,lg | 1.50 | 5.3 | 24 | |||
| 400,1x2,lg | 1.50 | 5.3 | 14 | |||
| 400,2x2,lg | 1.50 | 5.3 | 9 | |||
| NIR | 2k x 2k Hawaii 2RG [1] | NDR | 2.12 | 8-25 [2] | 0.88 | 45000 |
[2] The RON depend on the DIT and is larger for short DIT. The RON~8 corresponds to a DIT larger or equal than 300s.