GIRAFFE Low Resolution grating settings

SETUPS DEFINITION AND RESOLUTION.

Low resolution mode 600 lines/mm, 34.0 blaze grating.

Filter Theta Order p2pp name ETC name Wl start Wl center Wl end Band R IFU/Argus R MEDUSA
132.45L385.7LR1362.0385.7408.146.1128008000
227.94L427.2LR2396.4427.2456.760.3102006400
332.24L479.7LR3450.1479.7507.857.8120007500
426.33L543.1LR4501.5543.1583.181.696006000
530.63L614.2LR5574.1614.2652.478.3118007400
634.93L682.2LR6643.8682.2718.474.6137008600
724.72L773.4LR7710.2773.4834.3124.089005600
829.02L881.7LR8820.6881.7940.0119.0104006500
- Filter is the filter number;
- Theta is the angle of diffraction with the grating normal;
- Order is the working echelle order;
- Wl star, center, end are the wavelengths at the edges and center of the CCD;
- Band is the covered spectral passband;
- resolving power, R, obtained in IFU/Argus mode;
- resolving power, R, obtained in MEDUSA mode.